Image quality optimization of narrow detector dental computed tomography for paediatric patients
Danieli Moura Brasil1, Ruben Pauwels2,3,4, Wim Coucke5
11 Department of Oral Diagnosis, Division of Oral Radiology, Piracicaba Dental School, University of Campinas (UNICAMP) , Piracicaba, Sao Paulo , Brazil.
Objectives:
Dental CBCT exposure parameters should be optimized according to patient-specific indications, mainly for children that are most vulnerable to harmful effects of ionizing radiation. The aim of this study was to determine optimized kV settings for paediatric acquisitions for a dental CBCT device.
Methods:
Clinical and quantitative evaluations of image quality were performed using 5 and 10 years old (y/o) anthropomorphic phantoms. Technical evaluation was performed with the SEDENTEXCT-IQ phantom. Images were obtained using a PaX-i3D Green CBCT (Vatech, Korea) device, combining tube voltages ranging from 85 to 110 kV and 2 fields of view (FOVs: 21 × 19 and 12 × 9 cm), while maintaining the radiation dose fixed by adjusting the mA accordingly. Clinically, observers assessed images based on overall quality, sharpness, contrast, artefacts, and noise. For quantitative evaluation, mean grey value shift, % increase standard deviation, % beam-hardening and contrast-to-noise ratio were calculated. For technical evaluation, segmentation accuracy, contrast-to-noise ratio and full width at half maximum were measured. Biplot graphs were used to choose representative parameters, from which the best kV was selected for each protocol and evaluation. kV values that had no statistical differences (p > 0.05) with the best kV chosen were considered as having the same quality.
Results:
Clinically, 95 kV was found as a cut-off value. From the quantitative aspect, 85 kV (p < 0.05) showed the worst quality, except in 12 × 9 cm 5 y/o. Technically, 85 and 110 kV in the large FOV showed significantly worse quality for the large FOV.
Conclusion:
For paediatric indications, 95 kV or higher (and correspondingly low mA values) was found as optimal.
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Distribution of Stresses in a Narrow Rectangular Beam
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II