Effect of different osteotome surgical protocols on primary implant stability in poor quality bone: an ex vivo study
Anderson Jara Ferreira1,2, Andrés Cáceres-Barreno3, Hugo Gaêta-Araujo3
1Piracicaba School of Dentistry - UNICAMP, Av. Limeira 901-Areião, Piracicaba, São Paulo, 13414-903, Brazil. anderson_jara@hotmail.com.
Abstract:
To compare the effect of three types of surgical site preparation protocols on the primary stability of dental implants installed in poor quality bone. Pig tibia blocks were analyzed using X-ray microtomography to identify regions of poor bone quality like bone type IV, in which thirty osteotomies were performed using one of three techniques: conventional implant bed preparation (n = 10), undersized osteotomies (n = 10), and osteotomies prepared with expander-osteotomes (n = 10). The primary stability of the implants was evaluated using the final insertion torque and the implant stability quotient (ISQ). Statistical analysis included the Shapiro-Wilk test, One-way ANOVA, with Tukey's post-hoc and Pearson's correlation coefficient with a significance level of 5%. The undersized surgical technique obtained higher final insertion torque values (p < 0.05) and better results in terms of resonance frequency analysis values (p < 0.05). Pearson's correlation test showed a strong correlation between the variables (p < 0.05). According to the methodology used, the undersized surgical technique proved to be more advantageous in regions of low bone quality.