Underestimation of 68Ga PET/CT SUV caused by activity overestimation using default calibrator settings
Tom Sanderson1, James Solomon2, Chris Nottage2
1University College London Hospitals NHS Foundation Trust, United Kingdom.
Purpose:
A PET/CT scan of a uniformly filled 68Ga phantom resulted in an unexpectedly low SUVmean of 0.88. A potential contributing cause of underestimation of 68Ga SUV is overestimation of 68Ga activity in the Radionuclide Calibrator associated with the PET/CT scanner. To investigate this, a Radionuclide Calibrator cross-calibration exercise was performed.
Methods:
A source of 68Ga was measured in 5 Capintec CRC-55tR calibrators using the pre-set 68Ga calibrator factor of 416, and a Veenstra VDC-505 calibrator using a factor recommended by the manufacturer. The source was then measured in an externally located Fidelis Secondary Standard Radionuclide Calibrator. Manual adjustments were made to the Capintec calibrator factors to match the decay corrected Fidelis measurement, followed by a repeat PET/CT scan of a uniform 68Ga phantom.
Results:
The cross-calibration results showed that the 5 Capintec calibrators systematically overestimated 68Ga activity by 7.8-9.4% (mean 8.5%) compared to the Fidelis. The calibrator factors were adjusted to 456-464 to match the Fidelis measurement, and the repeat phantom scan resulted in a SUVmean of 0.97, within the local tolerance of 1.00 ± 5%. The result for the Veenstra calibrator was within the tolerance of ±5%.
Conclusions:
Underestimation of 68Ga SUV was primarily caused by overestimation of 68Ga activity using the pre-set calibrator factor setting on a Capintec CRC-55tR. Improvement in quantification accuracy was achieved by adjusting the 68Ga calibrator factor based upon a cross-calibration exercise. We recommend SUV checks using a uniform phantom and regular calibrator cross-calibration exercises for all isotopes used for quantitative PET/CT imaging.
More Related Videos
07:54Implantation and Monitoring by PET/CT of an Orthotopic Model of Human Pleural Mesothelioma in Athymic Mice
Published on: December 21, 2019
10:04Analysis of 18FDG PET/CT Imaging as a Tool for Studying Mycobacterium tuberculosis Infection and Treatment in Non-human Primates
Published on: September 5, 2017
Related Concept Videos
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Plotting and Calibrating the Root Locus
The maximum gain occurs at the breakaway points between open-loop poles on the real axis, while the minimum gain is...
Calibration Curves: Correlation Coefficient
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...
Setting Time of Cement