Multilevel Oblique Lumbar Interbody Fusion in Degenerative Lumbar Disc Disease with Instability
Published on: July 25, 2025
Biomechanical Effects of an Oblique Lumbar PEEK Cage and Posterior Augmentation
Hector Soriano-Baron1, Anna G U S Newcomb1, Devika Malhotra1
1Department of Neurosurgery, Barrow Neurological Institute, St. Joseph's Hospital and Medical Center, Phoenix, Arizona, USA.
Objective:
Lumbar interbody spacers are widely used in lumbar spinal fusion. The goal of this study is to analyze the biomechanics of a lumbar interbody spacer (Clydesdale Spinal System, Medtronic Sofamor Danek, Memphis, Tennessee, USA) inserted via oblique lumbar interbody fusion (OLIF) or direct lateral interbody fusion (DLIF) approaches, with and without posterior cortical screw and rod (CSR) or pedicle screw and rod (PSR) instrumentation.
Methods:
Lumbar human cadaveric specimens (L2-L5) underwent nondestructive flexibility testing in intact and instrumented conditions at L3-L4, including OLIF or DLIF, with and without CSR or PSR.
Results:
OLIF alone significantly reduced range of motion (ROM) in flexion-extension (P = 0.005) but not during lateral bending or axial rotation (P ≥ 0.63). OLIF alone reduced laxity in the lax zone (LZ) during flexion-extension (P < 0.001) but did not affect the LZ during lateral bending or axial rotation (P ≥ 0.14). The stiff zone (SZ) was unaffected in all directions (P ≥ 0.88). OLIF plus posterior instrumentation (cortical, pedicle, or hybrid) reduced the mean ROM in all directions of loading but only significantly so with PSR during lateral bending (P = 0.004), without affecting the compressive stiffness (P > 0.20). The compressive stiffness with the OLIF device without any posterior instrumentation did not differ from that of the intact condition (P = 0.97). In terms of ROM, LZ, or SZ, there were no differences between OLIF and DLIF as standalone devices or OLIF and DLIF with posterior instrumentation (CSR or PSR) (P > 0.5).
Conclusions:
OLIF alone significantly reduced mobility during flexion-extension while maintaining axial compressive stiffness compared with the intact condition. Adding posterior instrumentation to the interbody spacer increased the construct stability significantly, regardless of cage insertion trajectory or screw type.
Related Concept Videos
Stress on an Oblique Plane
The Thoracic Cage: Sternum
The sternum is the elongated bony structure on the anterior side of the thoracic cage. It consists of three parts: the manubrium, the body, and the xiphoid...
The Thoracic Cage: Ribs
Parts of a Typical Rib
A typical rib has a head, neck, and body. The posterior end of the rib is called the head, followed by a narrow neck. The head articulates primarily with the costal...
Cranial Bones: Superior and Posterior View
The frontal bone is the single bone that forms the forehead. At its anterior midline, between the eyebrows, there is a slight depression called the glabella. The frontal bone also forms the supraorbital margin of the orbit. Near the middle of this margin is the supraorbital foramen, the opening that provides passage for a sensory nerve to the forehead. The frontal bone is thickened just above each supraorbital margin,...
Buffer Effectiveness
The buffer capacity is the amount of acid or base that can be added to a given volume...
Framing Effects