Accuracy of blood glucose measurements using the NOVA StatStrip® glucometer during cardiac surgery: a prospective
Yosuke Nakadate1, Hiroaki Sato2, Patricia Roque3
1Department of Anesthesiology, University of Yamanashi, 1110 Shimokato, Chuo-city, Yamanashi, 409-3898, Japan. ynakadate@yamanashi.ac.jp.
Purpose:
The Nova StatStrip® Glucose Hospital Meter System (Nova Biomedical, Waltham, MA, USA) is United States Food and Drug Administration approved for point-of-care use in critically ill patients, but its use during cardiac surgery has not been evaluated. In this study, we compare glucose values obtained during cardiac surgery by StatStrip® with values obtained by a blood gas analyzer.
Methods:
Blood glucose concentrations were analyzed in 121 patients by the StatStrip point- of-care test (POCT) glucose monitor and the GEM® Premier™ 3000 blood gas analyzer (Instrumentation Laboratory Company, Bedford MA, USA). Arterial blood samples were taken at baseline (before surgery), before cardiopulmonary bypass (CPB), during early and late CPB, and 30 min after CPB. Accuracy of the StatStrip glucometer was analyzed using the Clinical and Laboratory Standards Institute (CLSI) POCT12-A3 criteria (criterion 1; 95% of samples should be ± 0.66 mMol·L-1 of reference glucose values < 5.5 mMol·L-1 and ± 12.5% for reference glucose values > 5.5 mMol·L-1, criterion 2; 98% of samples should be ± 0.83 mMol·L-1 of reference glucose values < 4.1 mMol·L-1 or 20% of the reference glucose for values > 4.1 mMol·L-1).
Results:
The accuracy of StatStrip glucose measurements at baseline (99%, 100%) and before CPB (95%, 98%), but not during (early: 84%, 97%; late: 83%, 96%) and after (92%, 100%) CPB, satisfied the CLSI POCT12-A3 criteria.
Conclusion:
Arterial blood glucose measurement by StatStrip was accurate before CPB, but lacked accuracy during and after CPB. Glucose values should be interpreted with caution when intensive glucose control protocols are being used during cardiac surgery.
Trial Registration:
ClinicalTrials.gov (NCT02729064); registered 5 April, 2016.
More Related Videos
06:27Detecting Establishment of Shared Blood Supply in Parabiotic Mice by Caudal Vein Glucose Injection
Published on: February 6, 2020
04:54Using a Combination of Indirect Calorimetry, Infrared Thermography, and Blood Glucose Levels to Measure Brown Adipose Tissue Thermogenesis in Humans
Published on: June 2, 2023
Related Concept Videos
Uncertainty in Measurement: Accuracy and Precision
Glucose Homeostasis: Regulation of Blood Glucose
During fasting, when blood glucose levels are low, the pancreas secretes glucagon. it...
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...
Hormones Regulating Blood Glucose
In addition to accelerating glucose uptake and utilization, insulin has...
Observational Studies
There are three types of observational studies – Prospective, retrospective, and cross-sectional.
Prospective Study
Prospective studies, also known as longitudinal or cohort studies, are carried out by collecting future data from groups sharing similar characteristics. One...
Blood Studies for Cardiovascular System II: CRP, Hcy, and Cardiac Natriuretic Peptide Markers
These markers indicate stress or strain on the heart muscle:
Natriuretic Peptides (BNP)
Cardiac myocytes produce these hormones in response to ventricular stretching...