Hemodynamic Comparison of Sutureless and Rapid-Deployment Valves with Conventional Bioprostheses
Max Gotzmann1, Manuel Wilbring2, Efstratios Charitos1
1Department of Cardiac Surgery, Mid-German Heart Center, Halle (Saale), Germany.
Background:
Sutureless and rapid-deployment bioprostheses (RPDs) are advantageous due to shorter aortic cross-clamp times (ACCs); however, few studies have investigated hemodynamic outcomes in comparison to conventional bioprostheses (CBs).
Methods:
All patients receiving the Perceval (LivaNova, London, United Kingdom) and Intuity (Edwards Lifesciences, Irvine, California, United States) valves from February 2016 to December 2017 were included (n = 61). For controls, patients who underwent aortic valve replacement using CB from 2015 to 2018 (n = 743) were propensity-matched (n = 108). Primary end points were mean gradient at discharge, true internal diameter (ID) after implantation, and paravalvular leakage. Secondary end points were ACCs, cardiopulmonary bypass times (CPBTs), mortality, and pacemaker implantation.
Results:
Age was 73.2 years (standard deviation [SD]: 7.6) in the RPD group and 72.9 years (SD: 7.3) in the CB group (p = 0.827). Median logistic EuroSCORE II was 3.4% (Q1: 2; IQ3: 5.3) and 3% (Q1: 1.9; IQ3: 5.6; p = 0.599). While ACCs and CPBT were shorter in RPDs (97 [SD: 31.4] vs. 125.4 minutes [SD: 62.1], p = 0.003; and 76.1 [SD: 25.7] vs. 89.7 minutes [SD: 34.3], p = 0.022), procedural times were similar (p = 0.257). True ID was 21.97 mm (SD: 1.79) in RPDs and 20.15 mm (SD: 1.70) in CB (p < 0.001).RPDs and CB resulted in comparable mean gradients (12.8 mm Hg [SD: 6.4] vs. 13.8 mm Hg [SD: 5.6]; p = 0.387) and rate of paravalvular regurgitation. There were no differences in the rates of pacemaker implantation and mortality.In a subanalysis of RPDs, mean gradient was 15.48 mm Hg (SD: 7.51) in Perceval (n = 21) and 10.79 mm Hg (SD: 4.78) in Intuity (n = 33; p = 0.010).
Conclusions:
RPDs provided comparable hemodynamic performance, although implanted valves were larger in true ID. Whether this provides superior hemodynamic performance during exercise and longer durability needs further investigation. Among RPDs, the Intuity valve provided lower mean gradients at rest than Perceval valve.
Related Concept Videos
Sign Convention
The normal force acts perpendicular to the beam's cross-section and can...
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Enolate Mechanism Conventions
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Evolutionary Relationships through Genome Comparisons