Related Experiment Video
Updated: Jan 27, 2026

Author Spotlight: Anterior HR-OCT as a Non-Invasive Tool for Characterizing Ocular Surface Squamous Neoplasia
Published on: August 9, 2024
Evaluation of anterior segment parameters with two anterior segment optical coherence tomography systems: Visante and
Dewang Angmo1, Reena Singh1, Shweta Chaurasia1
1Department of Ophthalmology, Glaucoma Research and Clinical Facility, Dr. Rajendra Prasad Centre for Ophthalmic Sciences, All India Institute of Medical Sciences, New Delhi, India.
Purpose:
To determine the comparability of anterior chamber biometric measurements in primary angle closure disease (PACD) patients using two commercially available anterior segment optical coherence tomography machines (ASOCT): Visante and Casia.
Methods:
This was a cross-sectional observational study, which included clinically, diagnosed cases of PACD. Anterior segment biometric measurements were done using Casia and Visante ASOCT. Parameters studied were central corneal thickness (CCT), anterior chamber depth (ACD), nasal (N) and temporal (T) angle opening distance at 500 μm (AOD500) and 750 μm (AOD750), and N and T trabecular iris space area at 500 μm (TISA500) and 750 μm (TISA750).
Results:
Total 36 PACD patients (72 eyes) with average age of 59.48 ± 7.95 years were recruited, out of which 25 were females (69.44%) and 11 males (30.56%). The mean measurements of CCT, ACD, AOD500, and TISA on Casia and Visante machines were 522.5 ± 34.75 μm and 539.55 ± 29.56 μm (P = 0.00); ACD- 2.144 ± 0.38 mm and 2.133 ± 0.39 mm (P = 0.487); AOD500-0.27 ± 0.16 μm and 0.21 ± 0.10 μm (P = 0.04); and TISA500-0.100 ± 0.07 μm and 0.063 ± 0.03 μm (P = 0.00), respectively. A statistically significant difference was noted in CCT, N and T AOD, and TISA. A good corelation for ACD and CCT (ACD = 0.9816 and CCT = 0.772) only were noted between the two machines. The Bland-Altman plot analysis of different parameters between two machines has revealed good agreement of measurement of ACD and CCT but poor agreement for rest of the parameters.
Conclusion:
It is advisable not use the two machines interchangeably because of the wide limits of agreement and poor correlation of angle measurement values of Casia and Visante ASOCT.
Related Concept Videos
Angle Closure Glaucoma: Treatment
Muscles of the Anterior Neck
Contact Angle
The adhesive force is the molecular force between molecules of different materials, that is, between the molecules of the solid and the liquid. The cohesive...
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Wave Parameters
Angle of Twist: Problem Solving

