Related Experiment Video
Updated: Jan 27, 2026

Amplifying and Quantifying HIV-1 RNA in HIV Infected Individuals with Viral Loads Below the Limit of Detection by Standard Clinical Assays
Published on: September 26, 2011
Vortex- and Centrifugation-Free Extraction of HIV-1 RNA
Rachel N Deraney1, Derek Troiano1, Richard Joseph2
1Center for Biomedical Engineering, School of Engineering, Brown University, 182 Hope Street, Providence, RI, 02912, USA.
Background And Objective:
HIV viral load measurements play a critical role in monitoring disease progression in those who are on antiretroviral treatment. In order to obtain an accurate measurement, rapid sample preparation techniques are required. There is an unmet need for HIV extraction instruments in resource-limited settings, where HIV prevalence is high. Therefore, the objective of our study was to develop a three-dimensional (3D) microfluidic system to extract HIV-1 RNA with minimal electricity and without complex laboratory instruments.
Methods:
A 3D microfluidic system was designed in which magnetic beads bound with nucleic acids move through immiscible oil-water interfaces to separate HIV-1 RNA from the sample. Polymerase chain reaction (PCR) amplification was used to quantify the total amount of HIV-1 RNA extracted as we optimized the system through chip design, bead type, carry-over volume, carrier RNA concentration, and elution buffer temperature. Additionally, the extraction efficiency of the 3D microfluidic system was evaluated by comparing with a Qiagen EZ1 Advanced XL instrument using 20 HIV-1-positive plasma samples.
Results:
Our method has near-perfect (100%) extraction efficiency in spiked serum samples with as little as 50 copies/mL starting sample. Furthermore, we report carry-over volumes of 0.31% ± 0.006% of total sample volume. Using the EZ1 Advanced XL as a gold standard, the average percentage HIV-1 RNA extracted using the microchip was observed to be 65.4% ± 24.6%.
Conclusions:
From a clinical perspective, the success of our method opens up its possible use in diagnostic tests for HIV in the remote areas where access to vortexes and centrifuges is not available. Here we present a proof-of-concept device which, with further development, could be used for sample preparation at the point of care.
More Related Videos
Related Concept Videos
Centrifugation
Centrifugal Force
Alternative RNA Splicing
There are five types of alternative RNA splicing that vary in the ways the pre-mRNA segments are removed or retained in the mature mRNA. The first...
RNA Stability
RNA Interference
This process occurs naturally in cells, often through the activity of genomically-encoded microRNAs. Researchers can take advantage of this mechanism by introducing synthetic RNAs to deactivate specific genes for research or therapeutic purposes. For example, RNAi could be used...
RNA Splicing

