Performance of a novel reusable pediatric pulse oximeter probe
Carina King1,2, Tisungane Mvalo3,4, Kristen Sessions3
1Institute for Global Health, University College London, London, UK.
Objective:
To assess the performance of reusable pulse oximeter probe and microprocessor box combinations, of varying price-points, in the context of a low-income pediatric setting.
Methods:
A prospective, randomized cross-over study comparing time to biologically plausible oxygen saturation (SpO2 ) between: (1) Lifebox LB-01 probe with Masimo Rad-87 box (L + M) and (2) a weight-appropriate reusable Masimo probe with Masimo Rad-87 box (M + M). A post hoc secondary analysis comparison with historical usability testing data with the Lifebox LB-01 probe and Lifebox V1.5 box (L + L) was also conducted. Participants, children aged 0 to 35 months, were recruited from pediatric wards and outpatient clinics in the central region of Malawi. The primary outcome was time taken to achieve a biologically plausible SpO 2 measurement, compared using t tests for equivalence.
Results:
We recruited 572 children. Plausible SpO2 measurements were obtained in less than 1 minute, 71%, 70%, and 63% for the M + M, L + M, and L + L combinations, respectively. A similar pattern was seen for less than 2 minutes, however, this effect disappeared at less than 5 minutes with 96%, 96%, and 95% plausible measurements. Using a ±10 second threshold for equivalence, we found L + M and M + M to be equivalent, but were under-powered to assess equivalence for L + L.
Conclusions:
The novel reusable pediatric Lifebox probe can achieve a quality SpO2 measurement within a pragmatic time range of weight-appropriate Masimo equivalent probes. Further research, which considers the cost of the devices, is needed to assess the added value of sophisticated motion tolerance software.
More Related Videos
Related Concept Videos
Pulse
The pulse serves as a clinical...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls...
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Regulation of Pulse
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...