Related Experiment Video
Updated: Jan 27, 2026

Monitoring Lung Function with Electrical Impedance Tomography in the Intensive Care Unit
Published on: September 6, 2024
Modelling approach to obtain regional respiratory mechanics using electrical impedance tomography and
1Electronics Engineering Department, Aeronautics Institute of Technology, São Paulo, Brazil.
Objective:
This paper presents a method for breath-by-breath estimation of regional respiratory mechanics without the need for special manoeuvres (such as inspiratory pause or low-flow inflation) using electrical impedance tomography (EIT) associated with pressure/airflow waveforms.
Approach:
We developed a method to estimate regional parameters using the regional impedance fraction, by multiplying it by global flow and volume waveforms. A volume-dependent elastance model was used to obtain compliance, resistance, volume-independent (E 1), and volume-dependent (E 2) components. Three swine under invasive mechanical ventilation were used to assess internal consistency and illustrate potential applications of our method. One animal (case 1) was ventilated with a broad range of tidal volumes to compare the consistency between regional and global resistances and compliances. Two other animals (cases 2 and 3) had respiratory compliance decreased, respectively, by overdistension and collapse as quantified by x-ray computed tomography.
Main Results:
In case 1, derived global estimates obtained from the independent regional estimates were strongly associated with direct measurements of global mechanics (correlation coefficients of 0.9976 and 0.9981 for compliances and resistances, respectively), suggesting consistency of our modelling. In cases 2 and 3, the development of lung overdistension and collapse over time was captured by regional estimates.
Conclusions:
Using EIT and pressure/airflow waveforms, regional respiratory parameters can be obtained cycle-by-cycle, refining lung function monitoring.
Significance:
The method allows real-time monitoring of regional parameters and their trends over time, which might be helpful to differentiate deterioration in lung compliance due to overdistension or collapse.
Related Concept Videos
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
The Quantum-Mechanical Model of an Atom
Respiratory Volumes and Capacities
Respiratory Volumes and Capacities I
Model Approaches for Pharmacokinetic Data: Physiological Models
Model Approaches for Pharmacokinetic Data: Compartment Models
Two primary types of compartment models are recognized: mammillary and catenary. The more...