Development and assessment of a multi-beam continuous-phantom-motion x-ray scatter projection imaging system
Christopher Dydula1, George Belev2, Paul C Johns1
1Department of Physics, Carleton University, 1125 Colonel By Drive, Ottawa, Ontario K1S 5B6, Canada.
Abstract:
X-ray image formation using scattered radiation can yield a superior contrast-to-noise ratio compared to conventional transmission x-ray imaging. A barrier to practical implementation of scatter imaging systems has been slow image acquisition. We have developed a projection imaging system which uses five monoenergetic pencil beams in combination with continuous phantom motion to achieve acquisition times that are practical for medical and security applications. The system was configured at the Canadian Light Source synchrotron and consists of a primary collimator, motorized stages for phantom translation, a flat-panel x-ray detector for measuring scattered x rays, and photodiodes for simultaneously measuring transmitted x rays. Image generation requires several corrections to raw data artifacts arising from the nature of the detector, x-ray source, and acquisition procedure. We developed a novel correction for pixel location inaccuracy arising from continuous phantom motion. A five-beam system had nearly five times faster acquisition than a single-beam system. Continuous motion acquisition was approximately 30 times faster than step-and-shoot acquisition. The total acquisition time for a 9 cm × 5 cm phantom with 8425 pixels was just over 2 min. Image quality was also assessed, in part to determine its relation to acquisition speed. The width of sharp material boundaries was found to be at a minimum equal to the pencil beam width (1.75 mm) and to have an additional width equal to the product of the phantom translation speed and the acquisition time per pixel (up to 1.0 mm in our experiments). Contrast-detail performance was independent of acquisition speed, depending only on phantom entrance x-ray fluence. Pixel signal-to-noise ratio measurements indicate that detector readout noise is important for the scatter data, even for phantom air kerma as high as 30 mGy. Images could be improved with a detector having lower readout noise and higher sensitivity. Its spatial resolution could be moderate. We confirmed that for the same range of λ-1 sin(θ/2), where λ is the x-ray wavelength and θ is the scattering angle, scatter images acquired using different beam energies (33-70 keV) had nearly identical contrast.
Related Concept Videos
X-ray Imaging
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Fischer Projections
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Newman Projections
The organic molecules rotate across the single bonds leading to numerous temporary three-dimensional structures of varying energy known as...