Johann-type laboratory-scale x-ray absorption spectrometer with versatile detection modes
Ari-Pekka Honkanen1, Sami Ollikkala1, Taru Ahopelto1
1Department of Physics, University of Helsinki, P.O. Box 64, FI-00014 Helsinki, Finland.
We developed an affordable lab X-ray absorption spectrometer for analyzing transition metals and actinides. This versatile instrument offers high energy resolution and multiple measurement modes for various applications.
Area of Science:
- Materials Science
- Analytical Chemistry
- Physics
Background:
- X-ray absorption spectroscopy (XAS) is crucial for characterizing elemental and chemical states.
- Existing laboratory XAS systems can be expensive and complex.
- There is a need for accessible, versatile XAS instrumentation for broader research applications.
Purpose of the Study:
- To present a cost-effective laboratory X-ray absorption spectrometer.
- To demonstrate its capabilities for analyzing K and L edges of various elements.
- To showcase its adaptability for diverse experimental setups and in situ studies.
Main Methods:
- Utilized a conventional X-ray tube source.
- Employed bent Johann-type crystal monochromators for energy selection.
- Designed for X-ray absorption spectroscopy in the 4-20 keV energy range.
- Incorporated transmission, fluorescence, and imaging measurement modes.
Main Results:
- Achieved typical energy resolution of 1-5 eV at 10 keV.
- Covered K edges of 3d transition metals and L edges of 5d transition metals and actinides.
- Demonstrated the instrument's modularity for accommodating complex sample environments.
- Presented a range of successful application examples.
Conclusions:
- The developed spectrometer provides a low-cost, high-performance solution for laboratory XAS.
- Its flexibility supports a wide array of research, including in situ investigations.
- This instrument democratizes access to advanced XAS capabilities for scientific research.
Related Concept Videos
IR Spectrometers
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
Mass Spectrometers
Types of Errors: Detection and Minimization
Absolute error in a measurement is the numerical difference from the true or central value. Relative error is the ratio between absolute error and the true or central value, expressed as a percentage.
Errors can be classified by source, magnitude, and sign. There are three types of errors: systematic, random, and gross.
Systematic or...
What is a Mode?
There can be more than one mode in a data set if multiple values have the same highest frequency. For instance, suppose that the Statistics exam scores of 20 students are: 50; 53; 59; 59; 63; 63; 72; 72; 72; 72; 72; 76; 78; 81; 83; 84; 84; 84; 90; 93. Here, the mode is 72, as it occurs most frequently, five times.
A data set with two modes is called bimodal. For example,...
Drug Absorption: Factors Affecting GI Absorption