Analytical simulator of proton radiography and tomography for different detector configurations
Chiara Gianoli1, Sebastian Meyer1, Lorena Magallanes2
1Faculty of Physics, Department of Medical Physics, Experimental Physics, Ludwig-Maximilians-Universität München, Munich, Germany.
Purpose:
An analytical simulator of ion Radiography (iRad) and Computed Tomography (iCT) for protons is proposed to serve as imaging benchmark for different detector configurations.
Methods:
The analytical simulator is applied to an anthropomorphic phantom and provides iRad and iCT benchmarks. Proton trajectories are traced relying on the Most Likely Path (MLP) algorithm. To simulate the proton trajectories the Multiple Coulomb Scattering (MCS) model embedded in the MLP algorithm is extended to non-uniform water equivalent materials according to variable substitution in the well-known statistical description in uniform water. The proton trajectories are instead estimated relying on the typical assumption of uniform water, thus causing intrinsic inaccuracies of the MLP algorithm. In this work the analytical simulator is used to explore and firstly compare the imaging performances of list-mode and integration-mode detector configurations with proton pencil beam scanning.
Results:
The intrinsic inaccuracies of the MLP algorithm affect the imaging performances of list-mode detector configuration, which nevertheless remains superior to integration-mode detector configuration for iCTs. For relatively higher proton statistics, comparable or better imaging performances are offered by integration-mode detector configuration for iRads (upto 29.2% of WET difference). Uncertainties of proton trajectories due to beam spot size are shown to compromise the imaging performances of integration-mode detector configuration, but also to affect the accuracy of the MLP algorithm for list-mode detector configuration.
Conclusions:
Based on MCS model in non-uniform water equivalent materials, the proposed simulation environment can serve for development and testing of dedicated imaging methodologies prior to and in combination with realistic Monte Carlo simulations.
More Related Videos
Related Concept Videos
Electron Configurations
The relative energies of the subshells determine the order in which atomic orbitals are filled (1s, 2s, 2p, 3s, 3p,...
Controller Configurations
Control-system compensation involves various configurations, most commonly series or cascade compensation, in which the controller...
Electron Configuration of Multielectron Atoms
Stability of Equilibrium Configuration
A stable equilibrium occurs when a system tends to return to its original position when given a small displacement, and the potential energy is at its minimum. An example of a stable equilibrium is when a cantilever beam is fixed at one end and a weight is attached to the other end. If the weight...
Configurations of BJT
The common base configuration is noted for its high voltage gain, positioning it as an ideal choice for single-stage amplifier circuits, such as microphone pre-amplifiers. A notable...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...