3-D printed photoreceptor phantoms for evaluating lateral resolution of adaptive optics imaging systems
Abstract:
With adaptive optics (AO), optical coherence tomography and scanning laser ophthalmoscopy imaging systems can resolve individual photoreceptor cells in living eyes, due to enhanced lateral spatial resolution. However, no standard test method exists for experimentally quantifying this parameter in ophthalmic AO imagers. Here, we present three-dimensional (3-D) printed phantoms, which enable the measurement of lateral resolution in an anatomically relevant manner. We used two-photon polymerization to fabricate two phantoms, which mimic the mosaic of cone photoreceptor outer segments at multiple retinal eccentricities. With these phantoms, we demonstrated that the resolution of two multimodal AO systems is similar to theoretical predictions, with some intriguing speckle effects.