[Expression of Motion Artifacts in OCT-Angiography Imaging in Healthy Subjects Using Two Different Devices]
Natasa Mihailovic1, Jost Lennart Lauermann1, Friederike Schubert1
1Klinik für Augenheilkunde, Universitätsklinikum Münster.
Background:
To compare the expression of motion artifacts in optical coherence tomography angiography (OCT-A) in healthy subjects using two different devices.
Methods:
In this study, 25 eyes of 25 healthy volunteers with no history of any ocular disease or ocular surgery were included. OCT-A imaging was performed using the RTVue XR Avanti (Optovue Inc., Fremont, California, USA) and the Spectralis OCT-A (Heidelberg Engineering, Heidelberg, Deutschland). The macula was imaged twice in each proband with active eye tracking (ET) using a 3 × 3 mm2 or a 10 × 10° scan, respectively. The expression of motion artifact was analyzed by two independent readers in the superficial OCT-angiogram using the Motion Artifact Score (MAS).
Results:
The signal strength index (SSI) was 73.0 ± 7.8 (Optovue) and 39.6 ± 3.6 (Heidelberg), which is equivalent to 73.0% (Optovue SSImax = 100 = 100%) and 79.2% (SSImax = 50 = 100%) of the maximum quality score. Both devices showed a very good image quality (mean MAS Optovue: 1.32 ± 0.551, mean MAS Heidelberg: 1.7 ± 0.789, p = 0.006). Of all measurements, quilting/banding was found in 20% of Optovue patients (10/50) and 6% of Heidelberg patients (3/50). Stretching was found in 4% of Optovue patients (2/50) and in 6% of Heidelberg patients (3/50). Vessel doubling was only seen in one Optovue angiogram (2%) as well as a displacement (2%). Blink lines only existed in three Heidelberg angiograms (6%).
Conclusion:
Despite different software and hardware approaches, both devices were able to take high-quality images with a very low prevalence of motion artifacts. Nevertheless, these artifacts still also occur in healthy subjects with good fixation. With regards to MAS, there was a high agreement between the two readers. However, the analysis of artifacts remains complex and requires experience as well as a precise assessment in evaluating OCT-A images.
Related Concept Videos
Bioavailability Study Design: Healthy Subjects Versus Patients
Equation of Motion: General Plane motion
Moreover, the body's center of mass experiences a rotational effect as a result of these couple moments. This rotation can be articulated as the...
Cable Subjected to Its Own Weight
A generalized loading function is employed to analyze a cable subjected to its own weight. This function considers the force acting along the cable's arc length rather than its projected length, providing a more accurate...
Absolute Motion Analysis- General Plane Motion
As the drone's propellers rotate, an upward force is generated that counteracts the force of gravity, enabling the drone to lift off from the ground. This initial movement of the drone is along a straight path, representing a form of translational motion. In this phase, every point on the...
Equation of Motion: General Plane motion - Problem Solving
The friction between the roller and the ground is characterized by two coefficients. The static friction coefficient is 0.15, while the kinetic friction coefficient is 0.1. These values are crucial in understanding the interaction between...
Projectile Motion: Example
When we speak of the range (R) of a projectile on...