Measurement precision and accuracy of high artificial enrichment 15 N and 13 C tracer samples
Neringa Mannerheim1, Roland A Werner1, Nina Buchmann1
1Institute of Agricultural Sciences, ETH Zurich, Universitätstrasse 2, 8092, Zurich, Switzerland.
Rationale:
Oversaturation of the Faraday cup amplifiers of isotope ratio mass spectrometers when using tracers that are highly enriched in heavier isotopes (up to 99.9%) remains a major bottleneck to obtaining high-precision measurements. The memory effect plays a key role in reducing tracer sample measurement precision and accuracy. Several sample preparation approaches are known to reduce memory effects and to improve tracer sample measurement precision. However, the potential benefits when using very high enrichment tracer samples (> +1000 mUr) have not been tested.
Methods:
In this study, we test how specific sample positioning for measurements and frequent use of natural isotope abundance reference materials within the sequence affects the precision and accuracy of isotopic ratio analyses when using a Flash elemental analyser coupled to a Deltaplus XP isotope ratio mass spectrometer for very high enrichment (> +22000 mUr) 15 N tracer sample measurements. Furthermore, we investigate if tracer sample dilution with natural isotope abundance materials reduces memory effects and increases measurement precision and accuracy when measurements of high-enrichment 15 N and 13 C biomass tracer samples are conducted.
Results:
Frequent use of natural isotope abundance materials and specific positioning increased 15 N tracer sample precision, but it had a negative effect on the precision of quality control substances. 15 N and 13 C tracer sample dilution improved measurement precision by a maximum of ±0.9 mUr; however, a strong linear relationship between the original and the calculated φ values was found. Highly enriched 15 N tracer samples caused a maximum memory effect of 0.11%. High levels of 15 N abundance within the samples affected measurement accuracy by an average of 6.7%.
Conclusions:
We conclude that highly enriched tracer samples do not require dilution before analysis. Tracer sample precision can be improved by using a specific measurement order of expected isotope abundance and by the frequent use of natural abundance reference materials.
More Related Videos
10:16Design and Operation of a Continuous 13C and 15N Labeling Chamber for Uniform or Differential, Metabolic and Structural, Plant Isotope Labeling
Published on: January 16, 2014
13:14Measuring Fluxes of Mineral Nutrients and Toxicants in Plants with Radioactive Tracers
Published on: August 22, 2014
Related Concept Videos
Uncertainty in Measurement: Accuracy and Precision
Accuracy and Precision
Improving Translational Accuracy
Improving Translational Accuracy
Other Nuclides: 31P, 19F, 15N NMR
While fluorine-19 and phosphorous-31 have high natural abundances (100%) and positive gyromagnetic ratios, nitrogen-15 has a low natural abundance and a negative gyromagnetic ratio. However, nitrogen-15 is still preferred over nitrogen-14 (which has a...
Accuracy, limits, and approximation
Accuracy is defined as the closeness of the measured value to the true or actual value. In engineering mechanics, repeated measurements are taken during theoretical or experimental analyses to ensure that the result is precise and accurate.
The accuracy of any solution is based on the...