Related Experiment Video
Updated: Jan 26, 2026

Tracking the Mammary Architectural Features and Detecting Breast Cancer with Magnetic Resonance Diffusion Tensor Imaging
Published on: December 15, 2014
Simultaneous Multislice Echo Planar Imaging for Accelerated Diffusion-Weighted Imaging of Malignant and Benign Breast
Sabine Ohlmeyer, Frederik Bernd Laun, Theresa Palm
1Siemens Healthcare, Erlangen, Germany.
Objectives:
Comparison of the diagnostic value of simultaneous multislice (SMS) accelerated diffusion-weighted echo planar imaging (EPI) of malignant and benign lesions of the breast compared with a reference EPI sequence.
Materials And Methods:
The study was approved by the institutional ethics committee. Sixty-eight patients were examined with a diffusion-weighted EPI (reference EPI; TE = 54 milliseconds; TR = 9000 milliseconds; TA, 3:27 minutes) and a diffusion-weighted SMS accelerated EPI (SMS EPI; acceleration factor 2; TE = 58 milliseconds; TR = 4300 milliseconds; TA, 1:53 minutes) in addition to the standard magnetic resonance imaging (MRI) protocol. Further acquisition parameters were as follows: 3 T MAGNETOM Skyra (Siemens Healthcare, Erlangen, Germany), 2.5-mm isotropic resolution, field of view = 185 to 190 × 350 mm, 62 slices, b = 50 and 800 s/mm with 1 and 4 averages, respectively. A dedicated 16-channel bilateral breast coil was used for imaging. Image quality was evaluated with respect to the presence of artifacts, signal voids, and quality of fat suppression. These parameters were rated using a 5-point Likert scale (1 = very strong to 5 = negligible). The apparent diffusion coefficient (ADC) was measured in 72 focal lesions (46 breast carcinomas and 26 benign lesions), and the diagnostic value of the 2 datasets was statistically evaluated and compared. The evaluation was performed a second time excluding cysts.
Results:
Artifacts and signal voids were negligible in both sequences (mean on Likert scale for reference EPI 4.68 vs SMS EPI 4.65, P = 0.52, and mean on Likert scale for reference EPI 4.85 vs SMS EPI 4.77, P = 0.14). Fat suppression was significantly better in SMS EPI (mean on Likert scale 3.28 vs 2.97, P < 0.001, Pearson r = 0.49). For benign lesions, the mean ADC in both EPI sequences was 1.86 · 10 mm/s. For malignant lesions, a mean ADC of 0.90 · 10 mm/s for the reference EPI and 0.89 · 10 mm/s for the SMS EPI was found. No significant difference between the EPI sequences was observed for ADC values (P = 0.75) and for the area under the curve (SMS, 0.985; no SMS, 0.975). The cutoff for differentiation of benign and malignant lesions was at ADC = 1.42 · 10 mm/s for SMS EPI (sensitivity, 1; specificity, 0.88) and at 1.23 · 10 mm/s for the reference EPI (sensitivity, 1; specificity, 0.92). Excluding the cysts, the cutoff for differentiation of benign and malignant lesions was at ADC = 1.11 · 10 mm/s for SMS EPI (sensitivity, 0.89; specificity, 0.93) and at 1.23 · 10 mm/s for the reference EPI (sensitivity, 1; specificity, 0.87).
Conclusions:
Our data indicate that SMS acceleration can be used for diffusion imaging in breast MRI in clinical practice. Simultaneous multislice EPI achieved the same diagnostic accuracy in breast MRI, but in a substantially reduced scan time.
Related Concept Videos
Echo
Imagine the sound is reflected back to the ears. Assuming that the source is very close to the human, the difference between hearing the two sounds—the emitted sound and the reflected sound—may be more than the minimum time for perceiving distinct sounds. If this is the case,...
Diffusion
Diffusion
Crystal Field Theory - Tetrahedral and Square Planar Complexes
Crystal field theory (CFT) is applicable to molecules in geometries other than octahedral. In octahedral complexes, the lobes of the dx2−y2 and dz2 orbitals point directly at the ligands. For tetrahedral complexes, the d orbitals remain in place, but with only four ligands located between the axes. None of the orbitals points directly at the tetrahedral ligands. However, the dx2−y2 and dz2 orbitals (along the Cartesian axes) overlap with the ligands less than the dxy,...
Weighted Mean
For example, consider the number of goals scored in the matches of a tournament. While computing the average number of goals scored in the tournament, it may be more important to...
Accelerators
The effectiveness of calcium chloride can...