Published on: April 11, 2025
Simultaneous Comparison of Electrocardiographic Imaging and Epicardial Contact Mapping in Structural Heart Disease
Adam J Graham1, Michele Orini1,2, Ernesto Zacur3
1Barts Heart Centre, Barts Health NHS Trust, London, United Kingdom (A.J.G., M.O., G.D., H.D., N.T.S., J.D.L., A. Cambridge, J.G., N.J.O., S.W.-A., M.L., M.F., M.J.E., A. Chow, S.S., M.D., R.J.S., R.J.H., P.D.L.).
Background:
The accuracy of ECG imaging (ECGI) in structural heart disease remains uncertain. This study aimed to provide a detailed comparison of ECGI and contact-mapping system (CARTO) electrograms.
Methods:
Simultaneous epicardial mapping using CARTO (Biosense-Webster, CA) and ECGI (CardioInsight) in 8 patients was performed to compare electrogram morphology, activation time (AT), and repolarization time (RT). Agreement between AT and RT from CARTO and ECGI was assessed using Pearson correlation coefficient, ρ AT and ρ RT, root mean square error, E AT and E RT, and Bland-Altman plots.
Results:
After geometric coregistration, 711 (439-905; median, first-third quartiles) ECGI and CARTO points were paired per patient. AT maps showed ρ AT=0.66 (0.53-0.73) and E AT=24 (21-32) ms, RT maps showed ρ RT=0.55 (0.41-0.71) and E RT=51 (38-70) ms. The median correlation coefficient measuring the morphological similarity between the unipolar electrograms was equal to 0.71 (0.65-0.74) for the entire signal, 0.67 (0.59-0.76) for QRS complexes, and 0.57 (0.35-0.76) for T waves. Local activation map correlation, ρ AT, was lower when default filters were used (0.60 (0.30-0.71), P=0.053). Small misalignment of the ECGI and CARTO geometries (below ±4 mm and ±4°) could introduce variations in the median ρ AT up to ±25%. Minimum distance between epicardial pacing sites and the region of earliest activation in ECGI was 13.2 (0.0-28.3) mm from 25 pacing sites with stimulation to QRS interval <40 ms.
Conclusions:
This simultaneous assessment demonstrates that ECGI maps activation and repolarization parameters with moderate accuracy. ECGI and contact electrogram correlation is sensitive to electrode apposition and geometric alignment. Further technological developments may improve spatial resolution.
Related Concept Videos
Ischemic Heart Disease: Overview
Atherosclerosis, the primary malefactor, orchestrates this dangerous condition. It manifests as the accumulation of fatty deposits, akin to insidious plaques, within arterial walls. As time elapses, these plaques metamorphose, hardening and...
Rheumatic Heart Disease I: Introduction
Contact-dependent Signaling
Gap Junctions
In animal cells, gap junctions are formed...
Rheumatic Heart Disease IV: Nursing Management
Rheumatic Heart Disease III: Medical Management
Contact Angle
The adhesive force is the molecular force between molecules of different materials, that is, between the molecules of the solid and the liquid. The cohesive...