System Accuracy Assessment of a Blood Glucose Meter With Wireless Internet Access Associated With Unusual
Andreas Pfützner1,2,3, Filiz Demircik1,2, Valeria Kirsch1,3
11 Pfützner Science & Health Institute, Mainz, Germany.
Background:
In recent randomized clinical trials, an unusual reporting pattern of glycemic data and hypoglycemic events potentially related to an internet enabled blood glucose meter (MyGlucoHealth, BGM) was observed. Therefore, this clinical study was conducted to evaluate the system accuracy of the BGM in accordance with the ISO15197:2015 guidelines with additional data collection.
Methods:
To investigate system accuracy, 10 of 3088 devices and 6 of 23 strip lots, used in the trials, were selected by a randomization procedure and a standard repeatability assessment. YSI 2300 STAT Plus was used as the standard reference method. The samples were distributed as per the ISO15197:2015 recommendations with 20 additional samples in the hypoglycemic range. Each sample was tested with 6 devices and 6 strip lots with double determinations.
Results:
Overall, 121 subjects with blood glucose values 26-423 mg/dL were analyzed, resulting in 1452 data points. In all, 186/1452 readings (12.8%) did not meet the ISO acceptance criteria. Data evaluated according to the FDA guidelines showed that 336/1452 (23.1%) readings did not meet the acceptance criteria. A clear bias toward elevated values was observed for BG <100 mg/dL (MARD: 11.0%).
Conclusions:
The results show that the BGM, although approved according to standard regulatory guidelines, did not meet the level of analytical accuracy required for clinical treatment decisions according to ISO 15197:2015 and FDA requirements. In general, caution should be exercised before selection of BGMs for patients and in clinical trials.
Related Concept Videos
Clinical Trials
There are four phases in a clinical trial. A phase one...
Clinical Trials: Overview
Glucose Homeostasis: Regulation of Blood Glucose
During fasting, when blood glucose levels are low, the pancreas secretes glucagon. it...
Unusual Results
According to the range rule of thumb, any value above or below two standard deviations, 2σ from the mean, μ is considered unusual.
Maximum unusual value =...
Hypoglycemia and Glucagon
Hormones Regulating Blood Glucose
In addition to accelerating glucose uptake and utilization, insulin has...