Optimized Management of Endovascular Treatment for Acute Ischemic Stroke
Published on: January 18, 2018
Force and aspiration on catheters utilized in the ADAPT technique in acute ischemic stroke: A bench top analysis
Mickey Smith1, Jonathan Pace1, Connie Ju2
1University Hospitals Cleveland Medical Center, The Neurological Institute, Department of Neurosurgery, 11100 Euclid Avenue, Cleveland, 44106 OH, USA; Case Western Reserve University School of Medicine, 2109 Adelbert Road, Cleveland, 44106 OH, USA.
Introduction:
Given the high morbidity and mortality of stroke, there remains a demand for techniques that provide rapid and safe intervention while improving time to recanalization. The direct aspiration first pass technique (ADAPT) uses force and aspiration for clot removal without the aid of separators or retrievers. In this study, we compare the force and aspiration qualities of commercially available catheters.
Methods:
Four different catheters with varying inner diameters were set up in a bench top model to test catheter tip pressure and flow rate. Catheter tip pressure was measured by attaching the catheter to a vacuum pressure gauge and an aspiration pump. The flow rate was calculated by measuring the volume of room temperature water aspirated through each catheter over a given time.
Results:
The Microvention Sofia catheter generated the greatest tip force (21.32 g), and the Stryker AXS Catalyst 6 catheter generated the smallest tip force (15.88 g). The Penumbra ACE 068 catheter and Medtronic ARC catheter measured 20.87 g and 16.78 g respectively. The ACE 068 had highest rate of aspiration at 289 mL/min, and the Catalyst 6 catheter had the lowest rate at 214. The Microvention Sofia catheter had the second highest rate while the ARC had the third highest rate, measuring 285 mL/min and 256 mL/min, respectively.
Conclusions:
When using the ADAPT technique, knowledge of the tip force and catheter flow rate of newer catheters with larger distal inner diameters may guide selection of aspiration catheters. While this study demonstrates differences in tip force and flow rate of different commercially available catheters, clinical translation will require further testing and evaluation.
Related Concept Videos
Intermolecular Forces
Electromotive Force
Intermolecular vs Intramolecular Forces
Intermolecular Forces in Solutions
When the strengths of the intermolecular forces of attraction between solute and solvent species in a solution are no different than those present in the separated components, the solution is formed with no accompanying energy change. Such a solution is called an ideal solution. A mixture of ideal gases (or gases such as helium and argon,...
Regulation of Stroke Volume
Preload refers to the degree of stretch on the heart before it contracts. It's analogous to the stretching of a rubber band; the more it's stretched, the more forcefully it snaps back. This concept is encapsulated in the Frank-Starling law of the...
Sample Preparation for Analysis: Advanced Techniques
Acid digestion with strong acids is commonly used to dissolve inorganic materials that are insoluble (do not dissolve) in water. This method can be useful for...