Comparison of Continuous Glucose Monitoring Accuracy Between Abdominal and Upper Arm Insertion Sites
Isabelle Isa Kristin Steineck1,2,3, Zeinab Mahmoudi2,4, Ajenthen Ranjan1,2,3,5
11 Department of Endocrinology, Copenhagen University Hospital Hvidovre, Hvidovre, Denmark.
Abstract:
The aim was to compare the accuracy of the Dexcom® G4 Platinum continuous glucose monitor (CGM) sensor inserted on the upper arm and the abdomen in adults. Fourteen adults with type 1 diabetes wore two CGMs, one placed on the upper arm and one placed on the abdomen. Three in-clinic visits of 5 h with YSI (2300 STAT, Yellow Springs Instrument) measurements as comparator were performed. Each visit was followed by 4 days with seven-point self-monitoring of blood glucose (SMBG) in free-living conditions. Accuracy analyses on the paired CGM-YSI and CGM-SMBG measurements of the two CGM sensors were performed. Using YSI as comparator, the overall Mean Absolute Relative Difference (MARD) for the CGMabd was 12.3% and CGMarm was 12.0%. The percentage of the CGM measurements in zone A of Clarke error grid analysis for the CGMabd was 85.6% and CGMarm was 86.0%. The hypoglycemia sensitivity for the CGMabd and CGMarm was 69.3%. Using SMBG as comparator, the overall MARD for the CGMabd was 12.5% and CGMarm was 12.0%. The percentage of the CGM measurements in zone A for the CGMabd was 84.1% and the CGMarm was 85.0%. The hypoglycemia sensitivity for the CGMabd was 60.0% and the CGMarm was 71.1%. All the P-values from the comparisons between the accuracy of CGMabd and CGMarm were >0.05. The accuracy of a Dexcom G4 Platinum CGM sensor placed on the upper arm was not different from the accuracy of the sensor placed on the abdomen in adults with type 1 diabetes.
More Related Videos
Related Concept Videos
Uncertainty in Measurement: Accuracy and Precision
Improving Translational Accuracy
Improving Translational Accuracy
Conserved Binding Sites
Binding sites are often located in large pockets, and if their location on a protein’s surface is unknown, it can be predicted using various approaches. The energetic method computationally...
Ligand Binding Sites
Protein-ligand interactions are quite specific; even though numerous potential ligands surround a cellular protein at any given time, only a particular ligand can bind to that protein. Moreover, a ligand binds only to a dedicated area on the surface of the protein, known as the...
Muscles that Move the Arm
The pectoralis major has two origins. Its clavicular head originates on the medial half of the clavicle. In contrast, the sternocostal head originates on the costal cartilages of ribs 1-6, the sternum, and the aponeurosis of the external oblique of the...