Validation of distal radius failure load predictions by homogenized- and micro-finite element analyses based on
A J Arias-Moreno1,2, H S Hosseini3, M Bevers1
1Orthopaedic Biomechanics, Department of Biomedical Engineering, Eindhoven University of Technology, De Zaale, Groene Loper 15, 5612AP, Eindhoven, The Netherlands.
Abstract:
This study developed a well-standardized and reproducible approach for micro-finite element (mFE) and homogenized-FE (hFE) analyses that can accurately predict the distal radius failure load using either mFE or hFE models when using the approaches and parameters developed in this study.
Introduction:
Micro-FE analyses based on high-resolution peripheral quantitative CT (HR-pQCT) images are frequently used to predict distal radius failure load. With the introduction of a second-generation HR-pQCT device, however, the default modelling approach no longer provides accurate results. The aim of this study was to develop a well-standardized and reproducible approach for mFE and hFE analyses that can provide precise and accurate results for distal radius failure load predictions based on second-generation HR-pQCT images.
Methods:
Second-generation HR-pQCT was used to scan the distal 20-mm section of 22 cadaver radii. The sections were excised and mechanically tested afterwards. For these sections, mFE and hFE models were made that were used to identify required material parameters by comparing predicted and measured results. Using these parameters, the models were cropped to represent the 10-mm region recommended for clinical studies to test their performance for failure load prediction.
Results:
After identification of material parameters, the measured failure load of the 20-mm segments was in good agreement with the results of mFE models (R2 = 0.969, slope = 1.035) and hFE models (R2 = 0.966, slope = 0.890). When the models were restricted to the clinical region, mFE still accurately predicted the measured failure load (R2 = 0.955, slope = 1.021), while hFE predictions were precise but tended to overpredict the failure load (R2 = 0.952, slope = 0.780).
Conclusions:
It was concluded that it is possible to accurately predict the distal radius failure load using either mFE or hFE models when using the approaches and parameters developed in this study.
Related Concept Videos
Reliability and Validity
Radius of Gyration of an Area
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Test for Homogeneity
Elements and Compounds
Elements
Elements are classified as atomic or molecular based on the nature of their basic units. They are unique forms of matter with specific chemical and physical properties that cannot break down into smaller substances by ordinary chemical reactions. There...
Imaging Studies for Cardiovascular System V: CT