Background:
Vitreoretinal surgeries require precise, dexterous, and steady instruments for operation in delicate parts of the eye. Robotics has presented solutions for many vitreoretinal surgical problems, but, in a few operations, the available tools are still not dexterous enough to carry out procedures with minimum trauma to patients. Vitrectomy is one of those procedures and requires some dexterous instruments to replace straight ones for better navigation to affected sides inside the eyeball.
Method:
In this paper, we propose a new vein puncturing solution with a 4-DOF motion to increase the workspace inside the eye. A two-member concentric tube-based 25G needle is proposed whose shape is optimized. To operate the concentric tube needle, a novel and miniaturized actuation system is proposed that uses hollow shaft motors for the first time. The presented prototype of actuation system has a stroke of 100 mm in a small size of 148 × 25 × 65 mm (L × W × H), suitable for approaching distant positions inside the eyeball.
Results:
Experimental results validate that the targeting accuracy of the needle is less than one millimeter and the needle tip can apply a force of 23.51 mN which is enough to perform puncturing. Furthermore, the proposed needle covers maximum workspace of around 128.5° inside the eyeball. For the actuation system, experiments show that it can produce repeatable motions with accuracy in submillimeter.
Conclusion:
The proposed needle system can navigate to the sites which are difficult to approach by currently available straight tools requiring reinsertions. Along with the miniaturized actuation system, this work is expected to improve the outcome of vitrectomy with safe and accurate navigation.
Related Concept Videos
PPE Use in Healthcare Settings II: Doffing
Concentration Cells
Consider the following voltaic cell:
Concentration and Rate Law
For example, in a generic reaction aA + bB ⟶ products, where a and b are stoichiometric coefficients, the rate law can be written as:
Uterine Tubes
Solution Concentration and Dilution
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....