A method to measure slice sensitivity profiles of CT images under low-contrast and high-noise conditions
Mitsunori Goto1, Chiaki Tominaga2, Masaaki Taura3
1Graduate School of Medicine, Tohoku University, Sendai 980-8575, Japan; Miyagi Cancer Center Natori, 981-1293, Japan.
Abstract:
Noise reduction features of iterative reconstruction (IR) methods in computed tomography might accompany the sacrifice of the longitudinal resolution, or slice sensitivity profile (SSP), at low contrast-to-noise ratio (CNR) conditions. To assess the benefit of IR methods correctly, the difference of SSP between IR methods and filtered-backprojection (FBP) must be taken into account. Therefore, SSP measurement under low-CNR conditions is necessary. Although edge methods are predominantly used, their performance under low-CNR conditions appears to be not fully established. We developed a method that is compatible with extremely low-CNR conditions. Thin plastic disk-shaped sheets embedded in acrylic resin were used as low-contrast test objects. The lowest peak contrast used was approximately 17 [HU]. We assessed the performance of our method by using FBP images. We identified a source of measurement instability aside from noise: the measured thin-slice SSP is dependent on the orbital phase of helical scan, presumably because of cone-beam artifacts. This impediment to high accuracy is manageable using phase-controlled scans. We confirmed that table position repeatability is much better than the value of the specifications, and therefore the ensemble-averaged images of multiple scans can be used for SSP measurement. Accurate measurement of SSP under extremely low-CNR conditions is possible, even when the test object is visually indiscernible from the noisy background. Low-contrast SSP behavior is elucidated for IR methods (AIDR-3D, FIRST, and AiSR-V) by using this measurement method.
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Measurement of Bioavailability: Pharmacokinetic Methods
Measurement of Bioavailability: Pharmacodynamic Methods