Breast intraoperative electron radiotherapy: Image-based setup verification and in-vivo dosimetry
Hamid Reza Baghani1, Mostafa Robatjazi2, Seied Rabi Mahdavi3
1Physics Department, Hakim Sabzevari University, Shohada-e Hastei Blvd, P.O. 9617976487, Sabzevar, Iran.
Introduction:
Single fraction nature of intraoperative radiotherapy highly demands a quality assurance procedure to qualify both beam setup and treatment delivery. The aim of this study is to evaluate the treatment setup during breast intraoperative electron radiotherapy (IOERT) and in-vivo dose delivery verification.
Materials And Methods:
Twenty-five breast cancer patients were enrolled and setup verification for each case was performed using C-arm imaging. The received dose by surface and distal end of target was measured by EBT2 film. The significance level of difference between obtained dosimetry results and predicted ones was evaluated by the T statistical test.
Results:
Acquired C-arm images in two different oblique views revealed any misalignment between the applicator and shielding disk. The mean difference between the measured surface dose and expected one was 1.8% ± 1.2 (p = 0.983) while a great disagreement, 11.1% ± 1.5 (p < 0.001), was observed between the measured distal end dose and expected one. This discrepancy is mainly correlated to the backscattering effect from the shielding disk. Target depth nonuniformities can also contribute to this remarkable difference.
Conclusion:
Employing the intraoperative imaging for IOERT setup verification can considerably improve the treatment quality. Therefore, it is suggested to implement this imaging procedure as a part of treatment quality assurance. Favorable agreement between the predicted and measured surface doses demonstrates the applicability of EBT2 film for dose delivery verification. The results of in-vivo dosimetry showed that the electron backscattering from employed shielding disk can affect the received dose by the distal end of tumor bed.
Related Concept Videos
Self-Evaluation: Self-Enhancement and Self-Verification
Strategies of Self-Presentation II: Self-Verification
Electron Carriers
Over the many stages of cellular respiration, glucose breaks down into carbon dioxide and water. Electron carriers pick up electrons lost by glucose in these reactions, temporarily storing and releasing them into the electron...
Electron Affinity
Electron Behavior
Electrons are negatively charged subatomic particles that are attracted to an orbit around the positively-charged nucleus of an atom. They reside in locations that are associated with energy levels called shells and are further organized into sub-shells and orbitals within each shell.
Electrons Orbit the Nucleus
Electrons are found in specific locations outside of the nucleus. The shell in which an electron resides indicates the general energy level of the electron: those closer to the...
Electron Transport Chains
The ETC is comprised of...