Multiple interleaved mode saturation (MIMOSA) for B1+ inhomogeneity mitigation in chemical exchange saturation
Andrzej Liebert1, Moritz Zaiss2, Rene Gumbrecht3
1Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Germany.
Purpose:
To mitigate inhomogeneity in quantitative CEST MRI at ultra-high magnetic field strengths (B0 ≥ 7 Tesla) using a parallel transmit system.
Methods:
Multiple interleaved mode saturation employs interleaving of 2 complementary phase sets during the saturation pulse train. Phase differences of 45° (first mode) and 90° (second mode) between 2 adjacent transmitter coil channels are used. The influence of the new saturation scheme on the CEST contrast was analyzed using Bloch-McConnell simulations. The presented method was verified in phantom and in vivo measurements of the healthy human brain. The relayed nuclear Overhauser effect was evaluated, and the inverse magnetic transfer ratio metric was calculated. Results were compared to a published correction method. All measurements were conducted on a whole-body 7 Tesla MRI system using an 8 transmitter and 32 receiver channel head coil.
Results:
Simulations showed that the inverse magnetic transfer ratio metric contrast of relayed nuclear Overhauser effect shows a smaller dependency on the relative amplitudes of the 2 different modes than the contrasts of Cr and amide proton transfer. Measurements of an egg white phantom showed markedly improved homogeneity compared to the uncorrected inverse magnetic transfer ratio metric (relayed nuclear Overhauser effect) images and slightly improved results compared to corrected images. In vivo multiple interleaved mode saturation images showed similar contrast compared to corrected images.
Conclusion:
Multiple interleaved mode saturation can be used as a simple method to mitigate inhomogeneity effects in CEST MRI at ultra-high magnetic field strengths. Compared to previous correction methods, acquisition time can be reduced because an additional scan, usually required for correction, can be omitted.
Related Concept Videos
Solution Equilibrium and Saturation
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Types of Chemical Reactions: Exchange and Reversible
A special kind of exchange reaction is the oxidation-reduction reaction, or the redox reaction. These reactions involve the transfer of electrons from one compound to another. The electrons in these reactions commonly come from hydrogen atoms, which consist of an electron and a proton. A molecule gives up a...
Energy Transfer in Chemical Reactions
Social Exchange Theory
What is a Mode?
There can be more than one mode in a data set if multiple values have the same highest frequency. For instance, suppose that the Statistics exam scores of 20 students are: 50; 53; 59; 59; 63; 63; 72; 72; 72; 72; 72; 76; 78; 81; 83; 84; 84; 84; 90; 93. Here, the mode is 72, as it occurs most frequently, five times.
A data set with two modes is called bimodal. For example,...