In vitro assessment of connection strength and stability of internal implant-abutment connections
Michelle D Kofron1, Matthew Carstens1, Cong Fu1
1Biomet 3i, LLC, a Zimmer Biomet company, 4555 Riverside Drive, Palm Beach Gardens, FL, United States.
Background:
Various connections have been machined to improve the fit between the dental abutment and implant. In vivo, the instability created by imprecisely fitting components can cause soft tissue irritation and bacterial colonization of the implant system. The aim of this study was to quantify abutment stability under in vitro force applications.
Methods:
Abutment stability and fit were quantitatively measured after application of rotational, vertical, and horizontal forces.
Findings:
The abutment connection held by friction (Friction-Fit) was the only group to have 0° angular rotation. A significantly greater vertical force was required to pull the abutment from the implant for the Friction-Fit connection as compared to all other experimental groups. The abutment connection held by a mechanically locking friction-fit with four grooves (CrossFit) and Friction-Fit demonstrated significantly lower lateral movement as compared to all other connections. The remaining connections evaluated included two hexagon connections that rely on screw placement for abutment fit (Conical + Hex #1 and Conical + Hex #2), one connection with protruding slots to align with recessed channels inside the implant (Conical + 6 Indexing Slots), and an internal connection that allows for abutment indexing every 120° (Internal Tri-Channel).
Interpretation:
Internal connection geometry influenced the degree of abutment movement. Friction-Fit and CrossFit connections exhibited the lowest rotational and horizontal motions. Significant differences were found between Friction-Fit and CrossFit following the application of a vertical force, with the Friction-Fit requiring a significantly greater pull force to separate the abutment from the implant.
Related Concept Videos
Dietary Connections
Functions of Connective Tissues
Hard connective tissues, such as bones and cartilage, provide structure and support to the body.
Loose Connective Tissue
Adipose Tissue
Adipose tissue consists primarily of fat storage cells called adipocytes and little extracellular matrix. A large number of capillaries present within adipose tissue allow rapid mobilization of lipid molecules. White adipose tissue is...
Introduction to Connective Tissues
Classification of Connective Tissues
Connective Tissue Proper
Connective tissue proper is the most abundant class of connective tissues. As its name implies, it predominantly connects different tissues in the body. Depending on the cell types, ground substance, viscosity, and fiber types in the ECM, connective tissue proper is further categorized into loose and dense....
Embryonic Connective Tissues
The mesenchyme is the first connective tissue that emerges in the developing embryo. It consists of loosely arranged multipotent mesenchymal cells and reticular fibers in the extracellular matrix. This loose arrangement allows easy migration of cells, which is essential for germ layer positioning, patterning, and organ morphogenesis during embryonic development.