Comparison between Roche Integra 400 plus and Abbott Architect ci8200 in Alanine aminotransferase assay
Somiya Lamrabat1,2, Amina Lyagoubi1,2, Ouardia Bouayadi1,2
1Central Laboratory of Mohammed VI University Hospital of Oujda City, Morocco.
Background:
The aim of this work is to present the results of a comparative study between the ALT assay on Integra 400 plus Roche Diagnostic versus Architect ci8200 of Abbott Diagnostic.
Methods:
A total of 200 patients hospitalized in the various departments of the university hospital Mohammed VI of Oujda were prospectively tested on two systems: Abbott Architect ci8200 and Roche Integra 400 plus. Both analyzers use the spectrophotometric technique by coupling the transamination reaction to the oxidation-reduction reaction at NAD. The agreement of the results between the different techniques was evaluated using the Bland-Altman difference diagram and the Passing-Bablok and Deming regression line.
Results:
There was a high concordance between the two assays: the equation of the Passing-Bablok line is YArchitect = -0,5625 + 0,9917 XIntegra with a correlation coefficient r2 = 0.999. The Bland-Altman diagram shows that the mean bias between the two methods is of the order of 1.1 IU/L and the difference between the ALT measurements by Architect and Integra is in the range of -1.4 to 3.6.
Conclusion:
Our study shows a high correlation of the ALT assay results between the architect ci8200 and Integra 400 plus.
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Evolutionary Relationships through Genome Comparisons
Comparison Between Electrical And Gravitational Forces
Since both are inverse square law forces, the distance gets canceled when the ratio of the two forces is considered. Instead, the ratio of the electrical and gravitational forces depends on...
Comparison between RL and RC circuits
Molecular Comparison of Gases, Liquids, and Solids