Visual performance of an extended depth of focus intraocular lens for treatment selection
Thomas Kohnen1, Myriam Böhm2, Eva Hemkeppler2
1Department of Ophthalmology, Goethe University, Frankfurt am Main, Germany. kohnen@em.uni-frankfurt.de.
Purpose:
To evaluate the visual performance of an extended depth of focus (EDOF) intraocular lens (IOL) for patient selection.
Methods:
This prospective non-comparative case series evaluated fifty-two eyes (26 patients) after bilateral implantation of the Tecnis Symfony IOL. Exclusion criteria were previous ocular surgeries, corneal astigmatism of >1.5 diopter, and ocular pathologies or corneal abnormalities. Uncorrected (UCVA) and distance-corrected (DCVA) visual acuity (VA) in 4 m, 80 cm, 60 cm and 40 cm (logMAR), defocus curve, contrast sensitivity (CS) and a questionnaire on subjective quality of vision (QoV) were assessed after 3 months.
Results:
UCVA was -0.02 ± 0.071 logMAR at 4 m, 0.01 ± 0.077 logMAR at 80 cm, 0.08 ± 0.187 logMAR at 60 cm and 0.22 ± 0.153 logMAR at 40 cm. Defocus curve testing showed a flat VA range from 0.00 to -1.50 D (-0.05-0.02 logMAR). At near distance from -2.0 to -2.5 D a dip of DCVA was observed (0.14-0.28 logMAR). CS under photopic and mesopic conditions without and with glare was 1.56 logCS, 0.86 logCS, 0.78 logCS, respectively. Spectacle independence was achieved in 100% for distance and intermediate vision and 71% of patients at near distance.
Conclusion:
This EDOF IOL provides excellent intermediate and far (<0.1 logMAR) and acceptable near UCVA (<0.3 logMAR). Defocus curve testing confirmed an EDOF between far and intermediate distance and showed a dip at near distance. It provides good QoV, CS, high spectacle independence, perception of few optical phenomena, and particularly good night-driving capabilities.
Related Concept Videos
Uniform Depth Channel Flow
Depth Perception and Spatial Vision
Uniform Depth Channel Flow: Problem Solving
What is Natural Selection?
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Gravitational Potential Energy for Extended Objects