Related Experiment Video
Updated: Jan 25, 2026

A Teleoperated Robotic System-Assisted Percutaneous Transiliac-Transsacral Screw Fixation Technique
Published on: January 6, 2023
Stent Screw-Assisted Internal Fixation (SAIF) of Severe Lytic Spinal Metastases: A Comparative Finite Element
Luigi La Barbera1, Alessandro Cianfoni2, Andrea Ferrari3
1Laboratory of Biological Structure Mechanics, Department of Chemistry, Materials and Chemical Engineering "G. Natta", Politecnico di Milano, Milan, Italy; Department of Mechanical Engineering, Polytechnique Montréal, Montréal, Québec, Canada; Sainte-Justine University Hospital Centre, Montréal, Québec, Canada.
Objective:
A new stent-screw-assisted internal fixation (SAIF) minimally invasive cement-augmentation technique has been introduced to treat patients with extreme osteolytic lesions of the vertebral body. The aim of the current finite element study, employing a spine model with an extreme osteolytic defect, was to assess the effect of the SAIF technique in reducing strains in the vertebral body in comparison with a standard surgical short posterior fixation.
Methods:
Different finite element models of a L1-S1 spine were developed, representing an intact condition (reference configuration), an extreme osteolysis condition, and its treatment, respectively with stand-alone SAIF, SAIF and posterior fixation, and with stand-alone posterior fixation. Each model was loaded to reproduce standing and upper body bending. Principal strains were calculated on the superior endplate, anterior and posterior cortical walls. A paired Wilcoxon test with a 0.05 significance level was performed to statistically analyze the results.
Results:
Median strains on the bony structures increased in the osteolysis model compared with the intact model, and the SAIF technique was effective in reducing such strains under both standing and flexion conditions. Additional posterior fixation, combined with the SAIF technique, produced minimal further reduction of the median strains on the bony structures. Stand-alone posterior fixation only shielded the osteolytic vertebra avoiding excessive displacements but failed in restoring the axial stiffness to values typical of the intact vertebra.
Conclusions:
The new SAIF technique resulted effective in restoring the load-bearing capacity of the extensively osteolytic vertebra; additional posterior fixation provided only further minor advantages.
Related Concept Videos
Lytic Cycle of Bacteriophages
Comparing the Survival Analysis of Two or More Groups
Internal Energy
Internal Receptors
Elements and Compounds
Elements
Elements are classified as atomic or molecular based on the nature of their basic units. They are unique forms of matter with specific chemical and physical properties that cannot break down into smaller substances by ordinary chemical reactions. There...
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.