Background:
High frequency optical coherence tomography (HF-OCT) is a novel intravascular imaging technology developed for use in the cerebral vasculature. We hypothesize that HF-OCT characterization of intrasaccular device neck coverage can prognosticate exclusion of the aneurysm from the circulation.
Methods:
Bifurcation and sidewall aneurysms were made in six dogs. Seven aneurysms were treated with next generation intrasaccular devices (NGID) and four with traditional platinum coils. HF-OCT was performed to interrogate gaps in the neck coverage, coil herniation, or acute thrombus formation. Animals were re-imaged at 7, 30, 90, and 180 days following aneurysm embolization. An automated image processing method segmented the devices at the neck of the aneurysm and quantified neck coverage. The largest coverage gap was used to predict aneurysm occlusion at 180 days.
Results:
No difference was found in occlusion rates between the coil and NGID groups (P=0.45). Successful segmentation of the NGID construct was achieved in all cases. A coverage gap >1 mm2 was found to predict failed aneurysm occlusion (P=0.047). This threshold was able to predict all cases of failed occlusion. The average number of devices needed to treat the aneurysm was lower in the NGID group (1.9 vs 6.75, P=0.009). HF-OCT showed strong agreement with scanning electron microscopy (bias 0.0024 mm2 (95% CI -0.0279, 0.0327)).
Conclusions:
HF-OCT enables precise and accurate measurement of coverage gaps at the neck of aneurysms treated with intrasaccular devices in vivo. We provide in vivo evidence that uniform aneurysm neck coverage by intrasaccular devices is critical for aneurysm occlusion.
Related Concept Videos
Ecological Succession
Test for Homogeneity
Predicting Molecular Geometry
Muscles of the Anterior Neck
Arteries of the Head and Neck
The internal carotid arteries supply blood to the anterior portion of the cerebrum. They enter the...
Veins of Head and Neck
On the other hand, the vertebral veins, unlike their arterial counterparts, are not primarily responsible for brain drainage. Instead, they drain the cervical vertebrae, spinal cord, and some small...