Related Experiment Video
Updated: Jan 25, 2026

Split Point Analysis and Uncertainty Quantification of Thermal-Optical Organic/Elemental Carbon Measurements
Published on: September 7, 2019
INTERCOMPARISON EXPERIMENT AND UNCERTAINTY ANALYSIS OF THREE MEASUREMENT METHODS FOR RADON PROGENY AT NIM
Yunxiang Wang1, Lei Zhang2, Zhijie Yang3
1State Key Laboratory of Nuclear Physics and Technology, School of Physics, Peking University, Beijing, China.
Abstract:
An intercomparison experiment of three radon progeny measurement methods, based on alpha spectrometry (Kerr method), alpha-beta spectrometry and liquid scintillation counter (LSC method), was carried out in the standard radon chamber at NIM. Both the consistency and the uncertainty analysis of three different methods were studied in detail. Results show that, at the EEC level of 4000 Bq/m3, the uncertainties of the alpha-beta spectrometry, LSC method and Kerr method were 3.78%, 6.41% and 9.46%, respectively, which are mainly contributed by detection efficiency uncertainty, measuring time uncertainty and counting statistical uncertainty in sequence. The comparison results at different concentration levels show that the measurement values of three methods are consistent within the error range of 3%. The alpha-beta spectrometry can be used as a reference standard method to achieve the simultaneous measurement of RaA, RaB, RaC concentrations and EEC in the radon chamber.
Related Concept Videos
Uncertainty in Measurement: Reading Instruments
Uncertainty in Measurement: Significant Figures
The Uncertainty Principle
Uncertainty in Measurement: Accuracy and Precision
Uncertainty: Overview
Uncertainty: Confidence Intervals