Diagnostic Performance of Carbon Monoxide Testing by Pulse Oximetry in the Emergency Department
Nuria Villalba1, Zachary T Osborn1, Pamela R Derickson1
1Department of Surgery, Larner College of Medicine, University of Vermont, Burlington, Vermont.
Background:
Carbon monoxide (CO) exposure causes roughly 40,000 emergency department (ED) visits annually and is commonly misdiagnosed. Whereas the standard method of carboxyhemoglobin (HbCO) measurement utilizes blood gas analysis, a noninvasive, FDA-cleared alternative exists. We evaluated the performance of pulse oximetry (SpCO) for identification of CO exposure in ED patients.
Methods:
We compared pulse oximetry to blood HbCO levels in a prospective observational study of adult and pediatric subjects recruited from the ED. Nurses screened a convenience sample of patients and referred those with SpCO ≥ 10% to research staff. Researchers also approached individuals who presented with signs and symptoms of CO toxicity. We determined diagnostic performance with a Bland-Altman analysis and calculated sensitivity and specificity for detection of elevated HbCO at thresholds of ≥ 10% and ≥ 15%. To optimize the potential sensitivity of SpCO for detection of CO toxicity, research technicians performed 3 SpCO readings within 5 min of the blood draw for laboratory measurement. A positive SpCO test was defined as any SpCO ≥ 10%.
Results:
42,000 patients were screened, 212 were evaluated, and 126 subjects were enrolled. Median HbCO level was 6% (range 1.6-21.9%). Limits of agreement were -10.3% and 8.1%. Of 23 individuals with elevated HbCO ≥ 10%, 13 were not suspected based on clinical assessment. Critically elevated HbCO was present in 6 individuals. Based on our a priori threshold of 10% for a positive test, pulse oximetry identified 14 of 23 subjects with HbCO ≥ 10%, with a sensitivity of 61% (95% CI 39-80%) and a specificity of 86% (95% CI 78-92%), and 5 of 6 subjects with HbCO ≥ 15%, with a sensitivity of 83% (95% CI 36-100%) and a specificity of 81% (95% CI 73-87%).
Conclusions:
Pulse oximetry underestimated HbCO and produced false negative results (ie, SpCO < 10% for all three measurements) in 17% of ED subjects with elevated HbCO ≥ 15%. Triage screening with pulse oximetry detected cases of elevated HbCO that were not suspected by the clinical provider.
More Related Videos
04:05Author Spotlight: Unveiling Prognostic Indicators in Heart Failure - The Role of Phase Angle and Bioelectrical Impedance Analysis
Published on: June 30, 2023
09:52Setting Up a Stroke Team Algorithm and Conducting Simulation-based Training in the Emergency Department - A Practical Guide
Published on: January 15, 2017
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Myasthenia Gravis: Diagnostic Tests
The edrophonium test is a diagnostic tool for myasthenia gravis. It involves...
Pericarditis II: Clinical Features and Diagnostic Tests
Atherosclerosis II: Clinical Manifestations and Diagnostic Tests
Myocarditis II: Clinical Features and Diagnostic Tests
Mitral Regurgitation II: Clinical Features and Diagnostic Tests