Evaluating the Performance of Cardiac Pulse Duplicators Through the Concept of Fidelity
Reynaldo A Rodriguez1, Kiran H Dellimore1, Jacobus H Müller2
1Biomedical Engineering Research Group, Mechanical & Mechatronic Engineering, Stellenbosch University, Private Bag X1, Matieland, 7602, South Africa.
Introduction:
The advanced design techniques used in modern prosthetic heart valve (PHV) development require accurate replication of the entire cardiac cycle. While cardiac pulse duplicator (CPD) design has a direct impact on the PHV test data generated, no clear guidelines exist to evaluate the CPD's performance. In response to this, we present a method to quantitatively assess CPD performance.
Materials And Methods:
A method to establish the fidelity of CPDs was formulated based on the pressure/time relationship and the error related to this relationship's target. This method was applied to assess the performance of a custom-made CPD. The performance evaluation included the assessment of the motion control system and overall repeatability of pressure measurements using a St Jude Epic 21 mm aortic valve.
Results:
The CPD's motion control system had an average root mean square error (RMSE) beat-to-beat tracking accuracy of 0.046 ± 0.008 mm. Assessment of the pressure measurements yielded a repeatability of < 2.4 ± 0.9 mmHg RMSE beat-to-beat differential pressure. The combination of pressure and its location within a heartbeat (fidelity) was within 5.0% of the individual targets for at least 95% of heartbeats.
Conclusion:
Fidelity can be used to objectively quantify the performance of various aspects of CPDs and to identify the cause of unexpected PHV or CPD behaviour. It also enables comparisons to be made among various CPDs in terms of overall performance. This approach may enable standardization of the assessment of CPD performance in the future.
More Related Videos
06:29Cardiac Magnetic Resonance for the Evaluation of Suspected Cardiac Thrombus: Conventional and Emerging Techniques
Published on: June 11, 2019
05:04In vitro Assessment of Cardiac Reprogramming by Measuring Cardiac Specific Calcium Flux with a GCaMP3 Reporter
Published on: February 22, 2022
Related Concept Videos
Pathophysiology of Cardiac Performance
Centrosome Duplication
To ensure that each daughter cell receives a centrosome after cell division, centrosome duplication...
Gene Duplication and Divergence
The duplicated copies of the gene are called Paralogs. Paralogs with similar sequences and functions form a gene family. Across several species, a large number of gene families are...
Duplication of Chromatin Structure
The basic unit of the chromatin is the nucleosome, consisting of DNA wrapped around octameric histone proteins and short stretches of linker DNA separating individual nucleosomes. The histone proteins within the nucleosome have their...
Self-Concept
Infancy and Emerging Recognition
During infancy, self-concept is virtually nonexistent. Babies do not distinguish themselves as separate entities and often mistake their...
Pulse
The pulse serves as a clinical...