Related Experiment Video
Updated: Jan 25, 2026

Serum and Plasma Copy Number Detection Using Real-time PCR
Published on: December 15, 2017
Bioanalytical method to detect midazolam in plasma.
Bádila Regina Dalla Costa1, Ana Paula Anzolin2, Luciana Grazziotin Rossato-Grando1,3
1Curso de Farmácia, Universidade de Passo Fundo, Brazil.
A new bioanalytical method accurately detects midazolam (MDZ) in plasma using HPLC-PDA and SPE. This validated technique offers precise monitoring for sedation levels, improving patient care.
Area of Science:
- Analytical Chemistry
- Pharmacology
Background:
- Midazolam (MDZ) is crucial for palliative sedation and anesthesia induction due to its rapid action.
- Current methods for monitoring patient sedation levels with MDZ lack accuracy.
- Accurate MDZ quantification in plasma is needed for improved clinical management.
Purpose of the Study:
- To develop and validate a sensitive bioanalytical method for detecting midazolam in human plasma.
- To establish a reliable tool for monitoring midazolam levels and patient sedation.
Main Methods:
- Developed a method using high-performance liquid chromatography (HPLC) with a photodiode array detector (PDA).
- Employed solid-phase extraction (SPE) for sample preparation.
- Validated the method for linearity, precision, accuracy, specificity, and sensitivity.
Main Results:
- The SPE method achieved a mean recovery of 101.03%.
- The method demonstrated linearity from 1.0-50.0 μg/mL with limits of detection (LD) and quantification (LQ) of 0.43 and 1.43 μg/mL, respectively.
- The method showed specificity for MDZ, with no interference from plasma anticoagulants.
Conclusions:
- The validated HPLC-PDA and SPE bioanalytical method is accurate, precise, and sensitive for midazolam detection in human plasma.
- This method provides a reliable approach for monitoring midazolam levels, potentially enhancing sedation management.
- The validated method supports future clinical applications in patient sedation monitoring.
Related Concept Videos
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
Precipitation Titration: Endpoint Detection Methods
In the Volhard method, a standard excess of AgNO3 is first added to the...
Composition of Blood Plasma
Plasma Membrane in Bacteria and Archaea
Enlargement of the Plasma Membrane
Protein Buffers in Blood Plasma and Cells
Certain amino acids can exist in a zwitterion state at a...