Adapted Surgical Procedure for Argus II Retinal Implantation: Feasibility, Safety, Efficiency, and Postoperative
Marie-Noëlle Delyfer1, David Gaucher2, Marc Govare3
1University of Bordeaux, Inserm, Bordeaux Population Health Research Center, team LEHA, UMR1219, Bordeaux, France; Department of Ophthalmology, Bordeaux University Hospital, Bordeaux, France.
Purpose:
To evaluate the feasibility, safety, and efficiency of an adapted surgical procedure used for postmarket Argus II implantations, so as to lower risks of postoperative hypotony or conjunctivoscleral erosion, and to describe the observed anatomic characteristics of the positioning of the implanted array.
Design:
Single-arm prospective multicenter clinical trial.
Participants:
Eighteen consecutive patients with end-stage retinitis pigmentosa.
Methods:
To protect the site of insertion of the cable of the device, a scleral flap was systematically added to the standardized implantation procedure. It was associated with temporalis fascia autograft, so as to cover the episcleral-fixed electronics case. Intraoperative and postoperative data at day 1, weeks 1 and 2, and months 1, 3, and 6 were collected. Postoperative distance between electrode-array and retina was measured on spectral-domain optical coherence tomography images. Position of the array was evaluated on fundus images between months 1 and 6.
Main Outcome Measures:
Feasibility of the modified surgical technique (time constraints, intraoperative complications), variations of intraocular pressure over time, postoperative ocular findings and adverse events, postoperative distance between the array and the retina, and rotation of the array between months 1 and 6 after implantation.
Results:
The adapted surgical technique was performed easily without associated specific complications. No cases of chronic hypotony or conjunctivoscleral erosion were reported. One serious device/procedure-related adverse event was recorded (sterile posterior uveitis), which resolved after vitrectomy. Postoperative distance between array and retina was variable: full apposition was achieved in 4 patients (22.22%), partial apposition observed in 9 patients (50.00%), and absence of strict apposition noted in 5 patients (27.78%, 4 of whom had posterior staphyloma). A statistically significant slight rotation of the array was observed between months 1 and 6 (P < 0.0001), occurring downwardly in 68.75% of cases.
Conclusions:
The combined use of scleral flap and temporalis fascia autograft was easily achieved and effective in preventing hypotony and conjunctival erosion in our study. Postoperative distance between semirigid array and retinal surface was variable, and increased in the case of preoperative staphyloma. A slight rotation of the device occurred over time. Further studies based on larger samples are needed to confirm our findings and determine their functional consequences.
More Related Videos
10:00Techniques for Processing Eyes Implanted with a Retinal Prosthesis for Localized Histopathological Analysis: Part 2 Epiretinal Implants with Retinal Tacks
Published on: February 14, 2015
12:01Techniques for Processing Eyes Implanted With a Retinal Prosthesis for Localized Histopathological Analysis
Published on: August 2, 2013
Related Concept Videos
Kidney Transplant II: Surgical Procedure
Survey Safety
Finding the Center of Gravity
Anatomical Positions
The body is upright, facing forward, and standing erect.
The feet are parallel and flat on the floor.
The arms are hanging by the...
Respiratory System Abnormal Finding I: Inspection and Percussion
Inspection Findings
During an inspection, several findings may suggest the presence of respiratory distress or disease. Pursed-lip breathing, where exhalation is slowed by...
Cardiovascular System Abnormal Findings I: Inspection and Palpation
Abnormal findings observed during an inspection