Related Experiment Video
Updated: Jan 25, 2026

Electrocardiogram Recordings in Anesthetized Mice using Lead II
Published on: June 20, 2020
Effect of lead design and pacing vector on electrical parameters of quadripolar coronary sinus leads: The RALLY-X4
Haran Burri1, Max Olaf Schrage2, Giovanni Morani3
1University Hospital of Geneva, Geneva, Switzerland.
Background:
Various lead designs have been developed to accommodate different coronary sinus anatomies. Our objectives were to compare electrical parameters of straight and spiral left ventricular leads, to evaluate capture thresholds and impedances using different pacing vectors, and to study evolution of thresholds over time.
Methods:
The RALLY-X4 study enrolled patients implanted with a lead from the Acuity X4 family (straight, spiral short, or spiral long). Electrical parameters (including capture thresholds from all 17 vectors) were measured at baseline and follow-up.
Results:
Data from 795 patients who were successfully implanted were analyzed. Straight and spiral leads had similar proportions of patients with thresholds <2.5 V/0.4 ms using the distal electrode (61-65% of patients) or from at least one of the proximal (E2-E4) electrodes (81-83% of patients). Unipolar vectors had significantly lower thresholds and impedances than bipolar vectors, with similar measurements compared to extended bipolar configurations. Capture thresholds increased with more proximal electrodes for all leads. Over a mean follow-up of 1 year, a slight decrease in capture thresholds was observed.
Conclusion:
Straight and spiral quadripolar leads allow to obtain clinically acceptable capture thresholds from at least one of the proximal electrodes in >80% of patients. Pacing vectors significantly affect electrical parameters, with higher thresholds in more proximal electrodes and lower thresholds with unipolar and extended bipolar configurations. Capture thresholds slightly decreased over a mean follow-up of 1 year.
Related Concept Videos
Phase-lead and Phase-lag Controllers
Time and frequency -Domain Interpretation of Phase-lead Control
The design of phase-lead control involves the strategic placement of poles and zeros to balance steady-state error and system...
Study Designs in Epidemiology
Observational studies are those where the researcher does not intervene but rather observes natural variations. They include cross-sectional, cohort, and...
Study Design in Statistics
Does aspirin reduce the risk of heart attacks? Is one brand of fertilizer more effective at growing roses than another? Is fatigue as dangerous to a driver as the influence of alcohol? Questions like these are answered using randomized experiments with proper...
Group Design
Acute Coronary Syndrome III: Diagnostic Studies