The Clinical Relationship of Anterior Capsular Tears and Vertical Spacing in the Femtosecond Laser Capsulotomy
Purpose:
To evaluate the frequency of anterior capsule tears and capsule edge irregularities in femtosecond laser anterior capsulotomy, including the effect of optimized vertical spacing settings.
Methods:
Three thousand eight hundred thirty-four cases were examined and grouped according to the femtosecond laser capsulotomy settings using manufacturer default settings with a vertical spacing of 10, 15, and 20 µm. Capsulotomy edge irregularities and anterior capsule tears were recorded intraoperatively.
Results:
The anterior capsule tear rates for the 10, 15, and 20 µm groups were 0.79%, 0.35%, and 0.09%, respectively. The capsule edge irregularity rates for the 10, 15, and 20 µm groups were 6.25%, 1.13%, and 1.15%, respectively. The treatment times for the 10, 15, and 20 µm groups were 1.6, 0.7, and 0.8 second, respectively. The risk reduction of anterior capsule tears positively correlated with the 15 and 20 µm group settings. The anterior capsule tear rate difference between the 10 and 20 µm group settings was statistically significant. Capsule edge irregularity rates were reduced in the 15 and 20 µm group settings. Eye movement was monitored in the 20 µm group and was observed in 9 of 13 patients with capsule edge irregularities.
Conclusions:
Anterior capsule tear rates are low with femtosecond laser cataract surgery. The rate can be further reduced with optimization of the settings, including an increase in vertical spacing. [J Refract Surg. 2019;35(5):280-284.].
More Related Videos
04:59Author Spotlight: Enhancing Visual Outcomes in Cataract Surgery: A Novel Technique to Prevent Posterior Capsular Opacification Through IOL Rotation
Published on: July 7, 2023
13:02Fabrication of 1-D Photonic Crystal Cavity on a Nanofiber Using Femtosecond Laser-induced Ablation
Published on: February 25, 2017
Related Concept Videos
Relationship Formation
Ending Relationships
Vertical Curve: Problem Solving
Introduction to Vertical Curves
State Space Representation
Consider an RLC circuit, a...
Space Trusses
At the core of a space truss lies the fundamental unit known as the tetrahedron. This structure is composed of six members that form a three-dimensional shape...