Experimental validations of a tunable-lens-based visual demonstrator of multifocal corrections
Vyas Akondi1,2, Lucie Sawides3, Yassine Marrakchi3
1Visual Optics and Biophotonics Laboratory, Instituto de Óptica "Daza de Valdés," Consejo Superior de Investigaciones Científicas (IO-CSIC), Madrid, Spain.
Abstract:
The Simultaneous Vision simulator (SimVis) is a visual demonstrator of multifocal lens designs for prospective intraocular lens replacement surgery patients and contact lens wearers. This programmable device employs a fast tunable lens and works on the principle of temporal multiplexing. The SimVis input signal is tailored to mimic the optical quality of the multifocal lens using the theoretical SimVis temporal profile, which is evaluated from the through-focus Visual Strehl ratio metric of the multifocal lens. In this paper, for the first time, focimeter-verified on-bench validations of multifocal simulations using SimVis are presented. Two steps are identified as being critical to accurate SimVis simulations. Firstly, a new iterative approach is presented that improves the accuracy of the theoretical SimVis temporal profile for three different multifocal intraocular lens designs - diffractive trifocal, refractive segmented bifocal, and refractive extended depth of focus, while retaining a low sampling. Secondly, a fast focimeter is used to measure the step response of the tunable lens, and the input signal is corrected to include the effects of the transient behavior of the tunable lens. It was found that the root-mean-square of the difference between the estimated through-focus Visual Strehl ratio of the multifocal lens and SimVis is not greater than 0.02 for all the tested multifocal designs.
Related Concept Videos
Distance Corrections
Reliability and Validity
Power Factor Correction
Data Validation
Nursing assessment guides are generally based on holistic models rather than medical...
Data Validation
Key parameters for method validation include:
NMR Spectrometers: Resolution and Error Correction