Early hemodynamic characteristics of eversion and patch carotid endarterectomies
Jesse Chait1, Michael Nicoara2, Pavel Kibrik2
1Division of Vascular Surgery, NYU Langone Hospital-Brooklyn, Brooklyn, NY, USA. jesse.chait@gmail.com.
Journal of Ultrasound
|May 10, 2019
Summary
This study found no significant hemodynamic differences between eversion and patch carotid endarterectomy (CEA) procedures. Post-operative peak systolic velocities were similar in both CEA techniques, indicating comparable early outcomes.
Area of Science:
- Vascular Surgery
- Cardiovascular Hemodynamics
- Medical Device Technology
Background:
- Carotid endarterectomy (CEA) is the standard surgical treatment for carotid artery stenosis.
- Eversion and patch CEA are common techniques, with studies showing similar clinical outcomes.
- Early hemodynamic differences between eversion and patch CEA techniques remain largely unknown.
Purpose of the Study:
- To investigate and compare early hemodynamic differences between eversion and patch carotid endarterectomy (CEA).
- To assess post-operative peak systolic velocities (PSV) in the internal carotid artery (ICA) and common carotid artery (CCA) following both procedures.
Main Methods:
- Retrospective analysis of 171 carotid endarterectomies (CEA) performed between March 2012 and June 2018.
- Data collected included patient demographics, procedure type, and pre- and post-operative duplex ultrasound (DUS) findings.
- Exclusion criteria included incomplete data, DUS >90 days post-op, concomitant bypass, isolated external carotid artery (ECA) endarterectomies, and re-do CEAs.
Main Results:
- 27 eversion and 81 patch CEAs were analyzed; patient demographics and laterality were similar, though eversion patients were older.
- Pre-operative peak systolic velocities (PSV) in the proximal ICA, distal ICA, and distal CCA were comparable between groups.
- Post-operative PSV and the change in PSV were not significantly different between eversion and patch CEA cohorts for any analyzed arterial segment.
Conclusions:
- Despite technical variations, eversion and patch CEA techniques result in similar early post-operative hemodynamics.
- No significant differences in post-operative PSV or PSV change were observed around the carotid bifurcation.
- These findings suggest comparable hemodynamic performance between the two CEA methods in the early post-operative period.
Keywords:
Carotid endarterectomyCarotid stenosisEversion endarterectomyHemodynamicsPatch endarterectomyUltrasoundVascular surgeryMore Related Videos
Related Concept Videos
Characteristics of Life
257.5K
Biology is a natural science that studies life and living organisms, including their structure, function, development, interactions, evolution, distribution, and taxonomy. The field's scope is extensive and divided into several specialized disciplines, such as anatomy, physiology, ethology, genetics, and many more. All living things share a few key traits, including cellular organization, heritable genetic material and the ability to adapt/evolve, metabolism to regulate energy needs, the...
257.5K
Long-patch Base Excision Repair
7.9K
Since the discovery of the two BER pathways, there has been a debate about how a cell chooses one pathway over the other and the factors determining this selection. Numerous in vitro experiments have pointed out multiple determinants for the sub-pathway selection. These are:
7.9K
Characteristics of Fluids
8.0K
When a force is applied parallel to the top surface of a solid, it resists the applied force due to the internal frictional forces between the layers of the solid known as shearing resistance. However, when the force is removed, the shearing forces restore the original shape of the solid. Other deformation forces also cause temporary changes in shape if the forces are not beyond a threshold magnitude. Solids tend to retain their shape, making the study of their rest and motion easier. Beyond...
8.0K
Characteristics of Fluids
992
Fluids differ from solids primarily in their molecular structure and stress response. Solids have tightly packed molecules with strong intermolecular forces, maintaining their shape and resisting deformation. In contrast, fluids have molecules spaced farther apart with weaker forces, allowing them to flow and deform easily.
Fluids, which include both liquids and gases, are substances that deform continuously under shearing stress. For example, water and oil are liquids with molecules that can...
Fluids, which include both liquids and gases, are substances that deform continuously under shearing stress. For example, water and oil are liquids with molecules that can...
992
Characteristics of BJT
1.3K
The Bipolar Junction Transistor (BJT), specifically in a common-emitter configuration, exhibits distinct current-voltage characteristics crucial for understanding its behavior in electronic circuits. These characteristics are established through experimental measurements of voltage and current relationships.
For input characteristics, the base-emitter voltage is varied, maintaining a constant collector-emitter voltage. This setup reveals a Shockley-type dependence of the collector current on...
For input characteristics, the base-emitter voltage is varied, maintaining a constant collector-emitter voltage. This setup reveals a Shockley-type dependence of the collector current on...
1.3K
Characteristics of JFET
1.2K
Junction Field Effect Transistors (JFETs) exhibit specific operational characteristics based on the relationship between the drain current (id) and the drain-source voltage (Vds), along with varying gate-source voltages (Vgs).
The core of a JFET's operation is controlling drain current by modulating the gate-source voltage. When the drain and gate voltage are set to zero, the JFET exhibits no net current flow, representing a state of equilibrium. The drain current increases linearly as the...
The core of a JFET's operation is controlling drain current by modulating the gate-source voltage. When the drain and gate voltage are set to zero, the JFET exhibits no net current flow, representing a state of equilibrium. The drain current increases linearly as the...
1.2K