Intra- and inter-operator reproducibility of US point shear-wave elastography in various organs: evaluation in
Riwa Kishimoto1, Katsuhiko Kikuchi2,3, Atsuhisa Koyama4
1National Institute of Radiological Sciences, National Institutes for Quantum and Radiological Science and Technology, 4-9-1 Anagawa, Chiba, 263-8555, Japan. kishimoto.riwa@qst.go.jp.
Purpose:
This study was conducted in order to assess the intra- and interoperator reproducibility of shear-wave speed (SWS) measurement on elasticity phantoms and healthy volunteers using ultrasound-based point shear-wave elastography.
Materials And Methods:
This study was approved by the institutional review board. Two operators measured the SWS of five elasticity phantoms and seven organs (thyroid, lymph node, muscle, spleen, kidney, pancreas, and liver) of 30 healthy volunteers with 1.0-4.5 MHz convex (4C1) and 4.0-9.0 MHz linear (9L4) transducers. The phantom measurements were repeated ten times, while the volunteer measurements were performed five times each. Intra- and interoperator reproducibility was assessed. Interoperator reproducibility was also evaluated with the 95% Bland-Altman limits of agreement (LOA).
Results:
In phantoms, all intraclass correlation coefficients (ICCs) were above 0.90 and the 95% LOA between the two operators were less than ± 18%. In volunteers, intraoperator ICCs were > 0.75 for all regions except the pancreas. Interoperator ICC was above 0.75 for the right lobe of the liver (depth 4 cm) and the kidney, but the 95% LOA was less than ± 25% only for the liver.
Conclusion:
Although excellent in phantoms, interoperator reproducibility was insufficient for all regions in the volunteers other than the right hepatic lobe at a depth of 4 cm. Clinicians should be aware of the 95% LOA when using SWS in patients.
Key Points:
• Our phantom study indicated a high reproducibility for shear-wave speed (SWS) measurements with point shear-wave elastography (pSWE). • In volunteers, intraoperator reproducibility was generally high, but the interoperator reproducibility was not high enough except for the right hepatic lobe at 4 cm depth. • To evaluate interoperator reproducibility, the 95% limits of agreement (LOA) between operators should be considered in addition to the intraclass correlation coefficient (ICC).
More Related Videos
05:56Implementation of Non-invasive Point of Care Transient Elastography for Evaluation of Liver Disease in Pediatric Populations with Cystic Fibrosis
Published on: August 29, 2025
09:02Cortical Bone Assessment Using Ultrasonic Guided Waves: A Reproducibility Study in a Healthy Population
Published on: January 31, 2025
Related Concept Videos
The Wave Nature of Light
Shearing Stress
The average shearing stress can be calculated by dividing the shear by the area of the cross-section.
Shearing Strain
Shear Diagram
First, a free-body diagram of the beam is drawn, representing all the external forces and internal reactions acting on the beam. One can calculate the reaction forces at each support by employing the equilibrium equations of force and moment. The vertical component...
Half wave rectifier
Full wave rectifier