Comparison of Tc-99m DMSA Renal Planar Scan and SPECT for Detection of Cortical Defects in Infants with Suspected
Ga Eun Kim1, Jae Hyun Park1, Joon Sik Kim1
1Department of Pediatrics, Keimyung University School of Medicine, Daegu, Republic of Korea.
Objective:
To compare the detection rate of Tc-99m dimercaptosuccinic acid (DMSA) renal single photon emission computerized tomography (SPECT) with that of planar scan for renal cortical defects in infants under one year of age with suspected acute pyelonephritis.
Methods:
Sixty infants under one year of age who exhibited fever and pyuria were prospectively included in this study. All infants underwent concurrent Tc-99m DMSA renal planar scan and SPECT. Detection rates for cortical defects and the numbers of cortical defects in both kidneys were compared between Tc-99m DMSA planar scan and SPECT. The cortical defects were classified as large and small; the rate of discordant results was compared between subgroups.
Results:
Negative findings by planar scan in 11 infants (11.3%) were modified to positive findings by SPECT. The detection rate of SPECT was significantly higher than that of planar scan (53.3% vs. 35.0%, p < 0.001). The numbers of cortical defects on SPECT were significantly higher than those on planar scan (0.85 ± 0.97 vs. 0.53 ± 0.83, p < 0.001). A significantly higher rate of discordant results between the planar scan and SPECT was observed for the small cortical defects compared to the large cortical defects (48.6% vs. 14.3%, p = 0.029).
Conclusions:
Tc-99m DMSA renal SPECT showed a higher detection rate, compared with planar scan, for renal cortical defects in infants with suspected acute pyelonephritis. The study results suggest that, compared with planar scan, Tc-99m DMSA SPECT can detect small cortical defects with greater sensitivity.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
06:38Early Detection of Drug-Induced Renal Hemodynamic Dysfunction Using Sonographic Technology in Rats
Published on: March 11, 2016
Related Concept Videos
Acute Pyelonephritis I: Introduction
Renal Drug Clearance: Comparison Between Renal Excretion Methods
Renal clearance is often associated with the renal glomerular filtration rate (GFR), which represents the rate at which plasma is filtered through the glomeruli in the kidney. When drug reabsorption is minimal and there is no active secretion, renal clearance is closely related to the...
Acute Pyelonephritis II: Diagnostic Studies and Management
Crystal Field Theory - Tetrahedral and Square Planar Complexes
Crystal field theory (CFT) is applicable to molecules in geometries other than octahedral. In octahedral complexes, the lobes of the dx2−y2 and dz2 orbitals point directly at the ligands. For tetrahedral complexes, the d orbitals remain in place, but with only four ligands located between the axes. None of the orbitals points directly at the tetrahedral ligands. However, the dx2−y2 and dz2 orbitals (along the Cartesian axes) overlap with the ligands less than the dxy,...
Leaky Scanning
The Sense of Self: Reflected Self-Appraisal and Social Comparison