High frame rate cardiac cine MRI for the evaluation of diastolic function and its direct correlation with
Jiming Zhang1, Jie Chen1, Benjamin Cheong1,2
1Department of Diagnostic and Interventional Radiology, CHI St. Luke's Health-Baylor St. Luke's Medical Center, Houston, Texas, USA.
Background:
Breath-hold cine MR is the method of choice for evaluating left ventricular (LV) systolic function; however, the evaluation of diastolic function remains in the domain of high frame rate echocardiography. Thus, a cine MR technique for simultaneously evaluating LV systolic and diastolic function would be clinically valuable.
Purpose:
To test the feasibility of extracting indices that characterize LV diastolic function from high frame rate cine MR.
Study Type:
Single center, prospective.
Population:
Asymptomatic volunteers (N = 24; age 45.8 ± 12.3 years).
Field Strength/Sequence:
High frame rate (70 fps) cine MR and phase-contrast MR during free breathing were acquired at 1.5T.
Assessment:
The following MR-based LV filling metrics were extracted from LV volume changes during the cardiac cycle: 1) the volume-rate ratio, REFP /RLFP (ratio of the peak LV filling rate during the early filling period [EFP] to that during the late filling period [LFP]); and 2) the volume ratio, VEFP /VLFP (the ratio of cumulative LV volume change between the EFP and LFP). These metrics were then compared with traditional transmitral blood flow-based MR and echocardiographic indices. The effect of temporal resolution on these metrics was also evaluated.
Statistical Tests:
Bland-Altman and linear regression analyses were used to evaluate the performance of the proposed metrics against traditional indices of diastolic function.
Results:
The REFP /RLFP and VEFP /VLFP correlated well with E/AQ-flow (r 2 = 0.66 and 0.54, respectively) and E/Aecho (r 2 = 0.58 and 0.49, respectively). Systolic indices remained robust (<3% error) for frame rates ≥20 fps. Although the proposed VEFP /VLFP was robust (<5% error) up to 25 fps, the proposed volume-rate diastolic function metrics were less reliable (>8% error) for frame rates below 35 fps.
Data Conclusion:
In asymptomatic volunteers, cardiac cine MR images acquired at frame rates >35 fps can be used to extract LV diastolic function indices from the temporal changes in LV volume.
Level Of Evidence:
1 Technical Efficacy Stage: 1 J. Magn. Reson. Imaging 2019;50:1571-1582.
Related Concept Videos
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
Framing Effects
Correlations
Frames
Frames are versatile and widely used in various applications such as structural supports for beams and columns, automobile chassis construction, and in the construction...
Correlation and Causation
Correlation versus Causation
If the dependent variable increases or decreases when the independent variable increases, there is a positive or negative...