Related Experiment Video
Updated: Jan 24, 2026

Ye's Swing Technique for Small-incision Lenticule Extraction Surgery
Published on: June 27, 2025
Decentration measurements using Placido corneal tangential curvature topography and Scheimpflug tomography pachymetry
Byunghoon Chung1, Hun Lee2, Cynthia J Roberts3
1Institute of Vision Research, Department of Ophthalmology, Yonsei University College of Medicine, Seoul, South Korea.
Purpose:
To assess the agreement of decentration measurements between tangential curvature topography and tomography pachymetry difference maps after a small-incision lenticule extraction (SMILE) procedure.
Setting:
Yonsei University College of Medicine and Eyereum Eye Clinic, Seoul, South Korea.
Design:
Retrospective case series.
Methods:
Decentrations and pupillary offset (angle κ) were analyzed and compared using two different methods. A difference map of the tangential curvature was generated with a Placido corneal topographer (Keratron Scout) and a pachymetry difference map was generated with a Scheimpflug tomographer (Pentacam) for each eye, using preoperative and 3-month postoperative scans. Decentrations were also compared using a vector difference plot for the two maps.
Results:
The study comprised 155 eyes (155 patients). The magnitude of decentration was 0.27 mm ± 0.17 (SD) (range 0.02 to 0.88 mm) with the tangential curvature difference maps and 0.26 ± 0.20 mm (range 0.00 to 0.91 mm) with the pachymetry difference maps, showing no significant difference in the arithmetic values of decentration (P = .124). However, the vector difference in pupillary offset between preoperative and postoperative measurements from the pachymetry difference map significantly correlated with the magnitude of vector differences of decentration between the two maps (P = .007). The vector difference plot also showed a wide distribution spread to all 4 quadrants, demonstrating a lack of agreement.
Conclusions:
Although there was no significant difference between the scalar magnitude of decentration with the tangential topography difference maps and tomography pachymetry difference maps, the vector differences did not agree and increasing error was present with increasing differences in pupillary offset in the pachymetry difference maps.
More Related Videos
10:21Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function
Published on: August 8, 2019
10:00Energy Dispersive X-ray Tomography for 3D Elemental Mapping of Individual Nanoparticles
Published on: July 5, 2016
Related Concept Videos
Degree of Curvature and Radius of Curvature
Methods of Obtaining Topography
Vertebral Column: Regions and Curvature
Regions of the Vertebral Column
In an adult, the spine is subdivided into five regions: the cervical, the thoracic, the lumbar, the sacral, and the coccygeal region. The spine initially develops as a series of 33 vertebrae; after 20 years of age, the nine bones in the sacral region, five sacral, and four coccygeal bones fuse to form...
Curvilinear Motion: Normal and Tangential Components
The positive direction of the t-axis aligns with the increasing position of the car along the curved path, denoted by the unit vector ut. Simultaneously, the n-axis, perpendicular to the t-axis, dissects the curved path into differential arc segments, each forming the arc of a circle with a radius of...
Space-Time Curvature and the General Theory of Relativity
This has been verified in many experiments. However, space and time are no longer absolute. Two observers moving relative to one another do not agree on the length of objects or the passage of time. The mechanics of objects based on Newton's laws of...
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...