Recommendation of New Medical Alarms Based on Audibility, Identifiability, and Detectability in a Randomized,
Christopher Bennett1, Roman Dudaryk2, Nichole Crenshaw3
1Music Engineering Technology, Frost School of Music, University of Miami, Coral Gables, FL.
Objectives:
Accurate and timely identification of existing audible medical alarms is not adequate in clinical settings. New alarms that are easily heard, quickly identifiable, and discernable from one another are indicated. The "auditory icons" (brief sounds that serve as metaphors for the events they represent) have been proposed as a replacement to the current international standard. The objective was to identify the best performing icons based on audibility and performance in a simulated clinical environment.
Design:
Three sets of icon alarms were designed using empirical methods. Subjects participated in a series of clinical simulation experiments that examined the audibility, identification accuracy, and response time of each of these icon alarms. A statistical model that combined the outcomes was used to rank the alarms in overall efficacy. We constructed the "best" and "worst" performing sets based on this ranking and prospectively validated these sets in a subsequent experiment with a new subject sample.
Setting:
Experiments were conducted in simulated ICU settings at the University of Miami.
Subjects:
Medical trainees were recruited from a convenience sample of nursing students and anesthesia residents at the institution.
Interventions:
In Experiment 1 (formative testing), subjects were exposed to one of the three sets of alarms; identical setting and instruments were used throughout. In Experiment 2 (summative testing), subjects were exposed to one of the two sets of alarms, assembled from the best and worst performing alarms from Experiment 1.
Measurements And Main Results:
For each alarm, we determined the minimum sound level to reach audibility threshold in the presence of background clinical noise, identification accuracy (percentage), and response time (seconds). We enrolled 123 medical trainees and professionals for participation (78 with < 6 yr of training). We identified the best performing icon alarms for each category, which matched or exceeded the other candidate alarms in identification accuracy and response time.
Conclusions:
We propose a set of eight auditory icon alarms that were selected through formative testing and validated through summative testing for adoption by relevant regulatory bodies and medical device manufacturers.
More Related Videos
10:52Simulation of Human-induced Vibrations Based on the Characterized In-field Pedestrian Behavior
Published on: April 13, 2016
07:31Author Spotlight: Advancing Cell Membrane Biophysics - Exploring Interactions and Challenges Through Experimental and Computational Approaches
Published on: September 1, 2023
Related Concept Videos
Bioequivalence Experimental Study Designs: Completely Randomized and Randomized Block Designs
Base Excision Repair
The first step of...
DNA Base Pairing
Lewis Acids and Bases
A coordinate covalent bond (or dative bond) occurs when one of the atoms in the bond provides both bonding electrons. For example, a coordinate covalent bond occurs when a water molecule combines with a hydrogen ion to form a hydronium ion. A coordinate covalent bond also results when...
Weak Base Solutions
Acids, Bases and Neutralization Reactions