Ligamentous Lisfranc Injury: A Biomechanical Comparison of Dorsal Plate Fixation and Transarticular Screws
Ankit Bansal1, Daniel A Carlson1, John R Owen1
1Orthopaedic Research Laboratory, Departments of Orthopaedic Surgery and Biomedical Engineering, Virginia Commonwealth University, Richmond, VA.
Objectives:
Optimal fixation technique after purely ligamentous Lisfranc injury remains controversial. This biomechanical study compares dorsal plate versus transarticular screw fixation by measuring dorsal and plantar joint diastasis. A unique protocol was developed, using reflective triad markers and positional cameras.
Methods:
Eleven cadaveric matched pairs were assigned to either transarticular screw or dorsal plate fixation. Two reflective triad markers were placed into the medial cuneiform (C1) and second metatarsal base (MT2). Three cameras recorded the 3-dimensional location of triads to quantify C1-MT2 diastasis in the following states: intact Lisfranc ligament (INTACT), cut ligament (CUT), fixed (SCREW or PLATE) joint, and fixed joint after 10,000 loaded cycles. On completion, the plantar Lisfranc ligament insertions were identified, and plantar diastasis was determined using additional reflective triads. Statistical post hoc pairwise comparisons assessed differences in diastasis.
Results:
C1-MT2 diastasis in the CUT state increased relative to INTACT (P < 0.001). SCREW fixation reduced C1-MT2 diastasis relative to CUT at dorsal (P < 0.007) and plantar (P = 0.015) locations after cycling. PLATE fixation significantly reduced dorsal diastasis relative to CUT (P < 0.001) but not for plantar diastasis (P > 0.99). PLATE plantar diastasis was numerically higher than INTACT but not significantly (P > 0.39). PLATE plantar diastasis tended to be greater than SCREW before cycling (P = 0.068) and after cycling (P = 0.080).
Conclusions:
Transection of the Lisfranc ligament complex yielded C1-MT2 diastasis. Both SCREW and PLATE fixation successfully reduced dorsal diastasis. However, upon load, the PLATE resulted in greater plantar diastasis, nearly statistically different relative to the SCREW. Cyclic loading at 343 N did not worsen diastasis.
Related Concept Videos
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Screw: Problem Solving
To evaluate the...
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
Fixation and Sectioning
The simplest type of preparation is the wet mount, in which the specimen is placed in a drop of liquid on the slide. A liquid specimen can be directly deposited on the slide using a dropper. Solid specimens, such as skin scraping, can be placed on the slide before adding a drop of liquid to prepare the wet mount. Sometimes the liquid is simply water, but stains are often added...
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...