Comparative Dose Accuracy of Durable and Patch Insulin Pumps Under Laboratory Conditions
Katharina Laubner1, Eva Singler1, Jan Straetener1
11 Division of Endocrinology and Diabetology, Department of Medicine II, Medical Center-University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany.
Abstract:
Recent studies demonstrate variable results of the accuracy with which patch pumps infuse insulin. Aim of this evaluation was to measure dose accuracies of the patch pump mylife™ OmniPod® (OP) in comparison with the durable insulin pump MiniMed® 640G (MM) simulating real-life clinical situations under laboratory conditions. Thirty-two OP and 15 MM were tested using insulin aspart at five different boluses (0.5, 1, 5, 10, and 15 international units [IU]) and three basal rates (0.2, 0.6, and 1.8 IU/h) at different time points during a 70 h investigation period. Owing to malfunctions only 22 OP and 11 MM could be analyzed. Dose accuracy was measured by an experimental setting based on IEC 60601-2-24:2012 with determination of weight differences of insulin collection tubes before and after experiments using a precision scale. A maximal tolerance of ±5% for boluses and basal rates was considered adequate according to IEC 60601-2-24:2012. For the five boluses, the percentages of measurement results within the ±5% accuracy threshold were as follows: OP (18.6%, 26.5%, 89.0%, 96.0%, and 96.0%); MM (21.7%, 44.1%, 88.1%, 98.3%, and 100.0%). Both pumps were more accurate at higher bolus volumes (5, 10, and 15 IU), later bolus periods, and if the accuracy threshold was lowered to <10%, <15%, or >15%. For the three basal rates, the percentages within the ±5% accuracy threshold were as follows: OP (66.7%, 22.7%, and 16.7%); MM (14.3%, 0.0%, and 0.0%). This study demonstrates low accuracy for basal rates and single bolus deliveries at low insulin doses for both pump models. Clinicians should be aware of this variability when initiating insulin pump therapy especially in insulin-sensitive patients with low insulin dose requirements.
More Related Videos
Related Concept Videos
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Uncertainty in Measurement: Accuracy and Precision
Improving Translational Accuracy
Brick Durability, Strength, and Appearance
ATP Driven Pumps III: V-type Pumps
The peripheral or cytosolic V1 domain with eight subunits is involved in ATP hydrolysis. The integral or transmembrane V0 domain containing at least five subunits...
ATP Driven Pumps II: P-type Pumps
A typical P-type pump has three cytosolic domains: nucleotide-binding (N), phosphorylation (P), and activator (A) domains. These domains are connected to the membrane-spanning helices by short amino acid segments. ATP hydrolysis and covalent phosphoenzyme intermediate formation are crucial parts of the catalytic cycle. At the highly...