Comparison of different registration methods and landmarks for image-guided radiation therapy of pulmonary tumors
Xiaohui Cao1, Ming Liu1, Fushan Zhai1
1Department of Radiotherapy and Oncology, The Third Hospital of Hebei Medical University, Shijiazhuang, 050051, Hebei, China.
Background:
To compare the accuracy, advantages and disadvantages of automatic registration methods at different anatomical-sites for thoracic image-guided radiation therapy (IGRT).
Methods:
The Varian-IX IGRT system was used to perform a manual registration of the images collected on the first fraction of 60 patients with lung cancer (42 cases central location and 18 cases of peripheral). The registered images were used as reference images. Offline registration was performed for computed tomography-CBCT images using four methods: whole image registration, ipsilateral registration, soft tissue tumor registration, and vertebral body registration. Time taken to complete and deviation value were analyzed between the different methods.
Results:
There were significant differences in absolute deviation value of all the three directions (P < 0.001) and the time consumption (P < 0.001) between 4 methods. The Z direction had significant differences in deviation value of 4 methods (0.023 ± 0.128 mm, - 0.030 ± 0.175 mm, - 0.010 ± 0.238 mm, - 0.075 ± 0.137 mm, P = 0.011). The difference was significant in the X direction of the ipsilateral registration method between central and peripheral lung cancer (0.033 ± 0.053 mm vs. 0.067 ± 0.067 mm, P = 0.045).
Conclusions:
The whole lung or affected side registration methods could be recommended to be used in the automatic registration function of the Varian-IX's On-Board Imaging (OBI) system.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Related Concept Videos
Biological Effects of Radiation
Renal Drug Clearance: Comparison Between Renal Excretion Methods
Renal clearance is often associated with the renal glomerular filtration rate (GFR), which represents the rate at which plasma is filtered through the glomeruli in the kidney. When drug reabsorption is minimal and there is no active secretion, renal clearance is closely related to the...
Physical Methods for Controlling Microbial Growth: Radiation and Filtration
Treatment for Pulmonary Arterial Hypertension: Oxygen Therapy for Respiratory Failure
Oxygen therapy is vital in increasing and maintaining blood oxygen levels in PAH patients. As a result, it aids in reducing fatigue,...
Radiation: Applications
The average...
The Sense of Self: Reflected Self-Appraisal and Social Comparison