The Practical Significance of Measurement Error in Pulmonary Function Testing Conducted in Research Settings
Richard B Belzer1, R Jeffrey Lewis2
1Mount Vernon, VA, USA.
Abstract:
Conventional spirometry produces measurement error by using repeatability criteria (RC) to discard acceptable data and terminating tests early when RC are met. These practices also implicitly assume that there is no variation across maneuvers within each test. This has implications for air pollution regulations that rely on pulmonary function tests to determine adverse effects or set standards. We perform a Monte Carlo simulation of 20,902 tests of forced expiratory volume in 1 second (FEV1 ), each with eight maneuvers, for an individual with empirically obtained, plausibly normal pulmonary function. Default coefficients of variation for inter- and intratest variability (3% and 6%, respectively) are employed. Measurement error is defined as the difference between results from the conventional protocol and an unconstrained, eight-maneuver alternative. In the default model, average measurement error is shown to be ∼5%. The minimum difference necessary for statistical significance at p < 0.05 for a before/after comparison is shown to be 16%. Meanwhile, the U.S. Environmental Protection Agency has deemed single-digit percentage decrements in FEV1 sufficient to justify more stringent national ambient air quality standards. Sensitivity analysis reveals that results are insensitive to intertest variability but highly sensitive to intratest variability. Halving the latter to 3% reduces measurement error by 55%. Increasing it to 9% or 12% increases measurement error by 65% or 125%, respectively. Within-day FEV1 differences ≤5% among normal subjects are believed to be clinically insignificant. Therefore, many differences reported as statistically significant are likely to be artifactual. Reliable data are needed to estimate intratest variability for the general population, subpopulations of interest, and research samples. Sensitive subpopulations (e.g., chronic obstructive pulmonary disease or COPD patients, asthmatics, children) are likely to have higher intratest variability, making it more difficult to derive valid statistical inferences about differences observed after treatment or exposure.
More Related Videos
Related Concept Videos
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.
PFTs involve using a spirometer, a...
Significance Testing: Overview
Errors In Hypothesis Tests
Accuracy and Errors in Hypothesis Testing
In hypothesis testing, the probability of making a Type I error, denoted as α, is commonly set at 0.05. This significance level indicates a 5%...
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Fundamental Attribution Error