A comparative study of NaI(Tl), CeBr3, and CZT for use in a real-time simultaneous nuclear and fluoroscopic
Wilco J C Koppert1, Martijn M A Dietze, Sandra van der Velden
1Radiology and Nuclear Medicine, Utrecht University and University Medical Center Utrecht, PO Box 85500, 3508 GA, Utrecht, The Netherlands.
Abstract:
Simultaneous acquisition of nuclear and fluoroscopic projections could be of benefit for image-guided radionuclide administration. A gamma camera positioned behind an x-ray flat panel detector can accomplish such simultaneous acquisition, but the gamma camera performance suffers from the intense x-ray dose. A regular NaI(Tl)-based camera has nominal performance up to 0.02 nGy dose per pulse, whereas 10 nGy dose is expected for our foreseen applications. We evaluated the performance of CeBr3- and CZT-based detectors and investigated a cost-effective improvement of a regular NaI(Tl)-based camera by the introduction of a high-pass filter and shorting circuit. A CeBr3-based detector was exposed to 5 mGy x-ray dose and the resulting light emission was measured over time to quantify the crystal afterglow, allowing comparison with a previously measured NaI(Tl)-based detector. The NaI(Tl)-, CeBr3- and CZT-based detectors were exposed to x-ray pulse sequences with dose from 0.06 to 60 nGy, while being irradiated with a gamma source. The mean gamma energy and energy resolution in between the x-ray pulses were measured as a reference of the detector performance. The afterglow signal after 3 ms was 14.1% for the NaI(Tl)-based detector, whereas for the CeBr3-based detector it was only 0.1%. The limits for a proper functioning detectors are 0.32 nGy for the NaI(Tl)-based detector with high-pass filter and shorting circuit and 18.94 nGy for the one with CeBr3. No energy degradation was observed for the CZT module in the studied dose range. The performance of regular NaI(Tl)-based gamma cameras deteriorates when exposed to high x-ray doses. CeBr3 and CZT are much better suited for introduction into a dual-layer detector but have high associated costs. Addition of a high-pass filter and shorting circuit into the PMT of a NaI(Tl)-based detector is a cost-effective solution that works well for low dose levels.
More Related Videos
04:36Author Spotlight: Real-Time Imaging of Bonding in 3D-Printed Layers
Published on: September 1, 2023
10:45Simultaneous Electroencephalography, Real-time Measurement of Lactate Concentration and Optogenetic Manipulation of Neuronal Activity in the Rodent Cerebral Cortex
Published on: December 19, 2012
Related Concept Videos
Nuclear Stability
To hold positively charged protons together...
Real Time RT-PCR
The real-time quantification of the number of amplified products is...
Nuclear Power
Nuclear Fuels
Nuclear fuel consists of a fissile isotope, such as uranium-235, which must be present in sufficient quantity to provide a...
Nuclear Fusion
A helium nucleus has a mass that is 0.7% less than that of four hydrogen nuclei; this lost mass is converted into energy during the fusion. This reaction produces about...
Non-nuclear Inheritance
Nuclear Binding Energy